Background: To study phacoemulsification probe shaft/wound and corneal surface/tip temperatures in vivo during longitudinal, torsional and combined phacoemulsification modes and their relationship to machine parameters.
Design: This was a prospective study at Chaithanya Eye Hospital and Research Institute, Trivandrum, India (tertiary).
Participants: Twenty-two eyes of 22 patients were randomized into six groups depending on the grade of nuclear sclerosis (NS) and the type of ultrasound used: Group 1, torsional in NS2; Group 2, torsional in NS3; Group 3, torsional with intermittent longitudinal in NS2; Group 4, torsional with intermittent longitudinal in NS3; Group 5, longitudinal in NS2; Group 6, longitudinal in NS3.
Methods: Patients underwent phacoemulsification by torsional, longitudinal or combined modalities. A thermal camera was used to measure phaco probe temperatures.
Main Outcome Measures: The mean probe shaft and tip temperatures were documented for different ultrasound modalities.
Results: The mean shaft and tip temperatures were: Group 1, 29.22°C±0.71°C and 28.4°C±0.88°C; Group 2, 32.12°C±0.62°C and 31.88°C±0.84°C; Group 3, 30.25°C±0.71°C and 29.35°C±0.62°C; Group 4, 31.95°C±0.65°C and 32.01°C±1.31°C; Group 5, 23°C and 27.6°C and Group 6, 23°C and 29.68°C, respectively. In all groups using longitudinal ultrasound, the phaco tip surface temperatures were higher than the shaft temperatures, except in Group 3. Shaft temperatures were higher than tip temperatures in cases using torsional phaco, except in Group 4. The mean temperature difference between groups was significant only for shaft temperatures (=0.001). On thermal imaging, for torsional phaco, the rise in temperature of the probe from shaft to tip was dependent on the amplitude of ultrasound applied (=0.009).
Conclusion: The shaft temperatures were higher than over the phaco tip during torsional phacoemulsification.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5291324 | PMC |
http://dx.doi.org/10.2147/OPTH.S123222 | DOI Listing |
Sci Rep
January 2025
The Fourth Engineering Co., LTD, China Railway Fourth Bureau, Hefei, 230012, People's Republic of China.
Research investigating the complex mechanical properties and energy evolution mechanisms of frozen calcareous clay under the influence of multiple factors is crucial for optimizing the artificial ground freezing method in shaft sinking, thereby enhancing construction quality and safety. In this study, a four-factor, four-level orthogonal test was devised, taking into account temperature, confining pressure, dry density, and water content. The complex nonlinear curvilinear relationship between deviatoric stress, volume strain, and axial strain of frozen calcareous clay under different interaction levels was analyzed.
View Article and Find Full Text PDFSci Rep
January 2025
School of Mechanical & Electronical Engineering, Lanzhou University of Technology, Lanzhou, 730050, China.
BMC Oral Health
January 2025
Department of Endodontics, School of Dentistry, Tehran University of Medical Sciences, Tehran, Iran.
Background: The fracture of an endodontic instrument within the root canal system can occur during root canal therapy, complicating thorough cleaning and shaping. Consequently, managing the broken fragment becomes crucial.
Methods: Eighty Nickel-titanium (NiTi) #20 K-files (Mani, Tochigi, Japan) were cut 8 mm from the tip, fixed into a corkboard, and classified into five groups (n = 14 each).
Sci Rep
January 2025
Jizhong Energy Fengfeng Group Co., Ltd, Handan, 056200, China.
Microsc Res Tech
December 2024
University of Science and Technology of China, Hefei, People's Republic of China.
Atomically resolved scanning tunneling microscope (STM) capable of in situ rotation in a narrow magnet bore has become a long-awaited but challenging technique in the field of strong correlation studies since it can introduce the orientation of the strong magnetic field as a control parameter. This article presents the design and functionality of a piezoelectrically driven rotatable STM (RSTM), operating within a 12 T cryogen-free magnet and achieving a base temperature below 1.8 K, along with spectroscopic capabilities.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!