Background: Accurate positioning of a ventricle catheter is of utmost importance. Various techniques to ensure optimal positioning have been described. Commonly, after catheter placement, additional manipulation is necessary to connect a burr hole reservoir or shunt components. This manipulation can lead to accidental catheter dislocation and should be avoided. Here, we present a new technique that allows direct endoscopic insertion of a burr hole reservoir with an already mounted ventricle catheter.

Methods: Before insertion, the ventricle catheter was slit at the tip, shortened to the correct length, and connected to the special burr hole reservoir. An intracatheter endoscope was then advanced through the reservoir and the connected catheter. This assemblage allowed using the endoscope as a stylet for shielded ventricular puncture. To confirm correct placement of the ventricle catheter, the endoscope was protruded a few millimeters beyond the catheter tip for inspection.

Results: The new technique was applied in 12 procedures. The modified burr hole reservoir was inserted for first-time ventriculoperitoneal shunting (n = 1), cerebrospinal fluid withdrawals and drug administration (n = 2), or different stenting procedures (n = 9). Optimal positioning of the catheter was achieved in 11 of 12 cases. No subcutaneous cerebrospinal fluid collection or fluid leakage through the wound occurred. No parenchymal damage or bleeding appeared.

Conclusions: The use of the intracatheter endoscope combined with the modified burr hole reservoir provides a sufficient technique for accurate and safe placement. Connecting the ventricle catheter to the reservoir before the insertion reduces later manipulation and accidental dislocation of the catheter.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.wneu.2017.01.095DOI Listing

Publication Analysis

Top Keywords

burr hole
24
hole reservoir
24
ventricle catheter
20
catheter
11
reservoir
8
catheter placement
8
optimal positioning
8
intracatheter endoscope
8
modified burr
8
cerebrospinal fluid
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!