Here we present two types of all-printable, highly stretchable, and inexpensive devices based on platinum (Pt)-decorated graphite for glucose determination in physiological fluids. Said devices are: a non-enzymatic sensor and an enzymatic biosensor, the latter showing promising results. Glucose has been quantified by measuring hydrogen peroxide (HO) reduction by chronoamperometry at -0.35V (vs pseudo-Ag/AgCl) using glucose oxidase immobilized on Pt-decorated graphite. The sensor performs well for the quantification of glucose in phosphate buffer solution (0.25M PBS, pH 7.0), with a linear range between 0 mM and 0.9mM, high sensitivity and selectivity, and a low limit of detection (LOD). Thus, it provides an alternative non-invasive and on-body quantification of glucose levels in human perspiration. This biosensor has been successfully applied on real human perspiration samples and results also show a significant correlation between glucose concentration in perspiration and glucose concentration in blood measured by a commercial glucose meter.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5328638 | PMC |
http://dx.doi.org/10.1016/j.bios.2017.01.058 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!