Introduction: One of the problems that the delivery of continuous positive airway pressure (CPAP) via a helmet poses is the generation of noise. The objective of our study was to assess the effect that the use of filter has on sound pressure levels generated by the delivery of positive airway pressure at different gas flow rates.

Materials And Methods: Sound pressure levels generated by neonatal helmet CPAP delivery were measured at different gas flows (20, 30, and 40 l/min) with and without a breathing filter. Noise intensity was measured by installing microphones in the inner ear of dummy heads wearing helmets.

Results: The sound pressure level increased by 38% at a gas flow of 40 l/min, as compared to a gas flow of 20 l/min {74 dBA [interquartile range (IQR) 2,2] vs 52 dBA (IQR 5,9), respectively}. Using the breathing filter as a diffuser has a variety of effects on sound pressure levels according to the gas flow rate.

Conclusion: The intensity of the noise generated by helmet delivery of positive airway pressure depends on the type of helmet used, gas flow, and use or not of a diffuser filter. Breathing filters with gas flows over 30 l/min might not be recommended since they would not attenuate but will rather amplify sound pressure.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5397998PMC
http://dx.doi.org/10.4103/1463-1741.199237DOI Listing

Publication Analysis

Top Keywords

sound pressure
20
gas flow
20
positive airway
16
airway pressure
16
pressure levels
12
pressure
9
noise generated
8
continuous positive
8
levels generated
8
delivery positive
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!