A simple and sensitive adsorptive stripping voltammetric procedure for the determination of trace concentration of U(VI) in natural water samples was developed. In order to remove surface active compounds from water sample solutions, a TiO/AlO photocatalysis system was applied. The linear calibration graph of U(VI) in the simultaneous presence of 2mgL anionic, cationic and nonionic surfactants, in the range from 5×10 to 2×10molL (180s) was achieved. The detection limit obtained in the solution after the use of UV-irradiation (10min) with TiO/AlO photocatalyst (0.9g) is equal to 2.3×10mol L. The presented procedure was successfully applied to uranium determination in SLEW-3 certified reference material, and to river water samples. In addition, a very low detection limit (2.9×10molL) for accumulation time of 180s was achieved due to application of a reversible deposited mediator (Zn) in the step of lead film electrode preparation.