The well-type ionization chamber has been designed for convenient use in brachytherapy source strength calibration. The chamber has a volume of 240 cm, weight of 2.5 kg, and is open to atmospheric conditions. The well-type ionization chamber dosimetric characteristics such as leakage current, stability, scattering effect, ion collection efficiency, and nominal response with energy were studied. The evaluated dosimetric characteristics of well-type ionization chamber were compared with two other commercially available well-type ionization chambers. The study shows that the newly developed well-type ionization chamber is reliable for air-kerma strength calibration. The results obtained confirm that this chamber can be used for the calibrations of high-dose rate brachytherapy sources.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5228046PMC
http://dx.doi.org/10.4103/0971-6203.195187DOI Listing

Publication Analysis

Top Keywords

well-type ionization
24
ionization chamber
20
newly developed
8
developed well-type
8
brachytherapy sources
8
strength calibration
8
dosimetric characteristics
8
chamber
7
well-type
6
ionization
6

Similar Publications

Article Synopsis
  • Thorium-227 was isolated from its decay products and its activity standardized using the TDCR method in liquid scintillation counting.
  • Additional measurements were conducted using efficiency tracing and live-timed anticoincidence counting to confirm the results.
  • The effectiveness of the separation process was validated through gamma-ray spectrometry, and various calibration methods for different detection devices are discussed.
View Article and Find Full Text PDF

Absolute gamma-ray emission intensities for 36 characteristic gamma rays from the decay of Ra, Pb, and their progeny were determined by measuring sources calibrated for activity by means of primary methods based on well-defined high-purity germanium (HPGe) detectors at both NIST and NPL. Results from the two laboratories agree with recent data evaluations, except for gamma rays with low emission intensities. The decay schemes have been re-balanced based on the new results.

View Article and Find Full Text PDF
Article Synopsis
  • This technical report discusses the quality control status of iodine (I) seed source strength measurement for Permanent Prostate Brachytherapy in Japan, highlighting issues with traceability in source strength assessments.
  • A working group was formed by JASTRO in 2021 to tackle the challenges faced by medical facilities, including a lack of consistent source strength verification, with a survey revealing that 41% of facilities do not confirm the number of seeds or measure source strength.
  • The report emphasizes the importance of the single-seed assay as a reliable method for ensuring traceability, noting that most facilities in Japan do not utilize this standardized measurement technique.
View Article and Find Full Text PDF

Dosimetric analysis of intracavitary brachytherapy applicators: a practical study.

Radiat Oncol J

September 2022

Department of Radiological Physics and Bio-engineering, Sher-I-Kashmir Institute of Medical Sciences, Jammu and Kashmir, India.

Purpose: Intracavitary brachytherapy is one of the important methods of gynecological cancer treatment. The effect of attenuation is not considered in the dose calculation method released by the American Association of Physicists in Medicine (AAPM) Task Group No. 43 Report (TG-43).

View Article and Find Full Text PDF

Background: Electronic brachytherapy (eBT) is considered a safe treatment with good outcomes. However, eBT lacks standardized and independent dose verification, which could impede future use.

Purpose: To validate the 3D dose-to-water distribution of an electronic brachytherapy (eBT) source using a small-volume plastic scintillation detector (PSD).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!