Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Background: We developed and validated a kinetic microplate hemolytic assay (HA) to quantify classical and alternative complement activity in a single dilution of human plasma or serum.
Methods: The assay is based on monitoring hemolysis of sensitized sheep (or uncoated rabbit) red blood cells by means of a 96-well microplate reader. The activity of the calibrator was evaluated by reference to 200 healthy adults. The conversion of 50% hemolysis time into a percentage of activity was obtained using a calibration curve plotted daily.
Results: The linearity of the assay as well as interference (by hemolysis, bilrubinemia and lipemia) was assessed for classical pathway (CP). The within-day and the between-day precision was satisfactory regarding the performance of commercially available liposome immunoassay (LIA) and ELISA. Patients with hereditary or acquired complement deficiencies were detected (activity was measured <30%). We also provided a reference range obtained from 200 blood donors. The agreement of CP evaluated on samples from 48 patients was 94% with LIA and 87.5% with ELISA. The sensitivity of our assay was better than that of LIA, and the cost was lower than either LIA or ELISA. In addition, this assay was less time consuming than previously reported HAs.
Conclusions: This assay allows the simultaneous measurement of 36 samples in duplicate per run of a 96-well plate. The use of a daily calibration curve allows standardization of the method and leads to good reproducibility. The same technique was also adapted for the quantification of alternative pathway (AP) activity.