Background: The process of assuring the safety of medical devices is constrained by reliance on voluntary reporting of adverse events. We evaluated a strategy of prospective, active surveillance of a national clinical registry to monitor the safety of an implantable vascular-closure device that had a suspected association with increased adverse events after percutaneous coronary intervention (PCI).
Methods: We used an integrated clinical-data surveillance system to conduct a prospective, propensity-matched analysis of the safety of the Mynx vascular-closure device, as compared with alternative approved vascular-closure devices, with data from the CathPCI Registry of the National Cardiovascular Data Registry. The primary outcome was any vascular complication, which was a composite of access-site bleeding, access-site hematoma, retroperitoneal bleeding, or any vascular complication requiring intervention. Secondary safety end points were access-site bleeding requiring treatment and postprocedural blood transfusion.
Results: We analyzed data from 73,124 patients who had received Mynx devices after PCI procedures with femoral access from January 1, 2011, to September 30, 2013. The Mynx device was associated with a significantly greater risk of any vascular complication than were alternative vascular-closure devices (absolute risk, 1.2% vs. 0.8%; relative risk, 1.59; 95% confidence interval [CI], 1.42 to 1.78; P<0.001); there was also a significantly greater risk of access-site bleeding (absolute risk, 0.4% vs. 0.3%; relative risk, 1.34; 95% CI, 1.10 to 1.62; P=0.001) and transfusion (absolute risk, 1.8% vs. 1.5%; relative risk, 1.23; 95% CI, 1.13 to 1.34; P<0.001). The initial alerts occurred within the first 12 months of monitoring. Relative risks were greater in three prespecified high-risk subgroups: patients with diabetes, those 70 years of age or older, and women. All safety alerts were confirmed in an independent sample of 48,992 patients from April 1, 2014, to September 30, 2015.
Conclusions: A strategy of prospective, active surveillance of a clinical registry rapidly identified potential safety signals among recipients of an implantable vascular-closure device, with initial alerts occurring within the first 12 months of monitoring. (Funded by the Food and Drug Administration and others.).
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6145132 | PMC |
http://dx.doi.org/10.1056/NEJMoa1516333 | DOI Listing |
J Vasc Access
January 2025
Division of Thoracic and Cardiovascular Surgery, Department of Surgery, Kaohsiung Chang Gung Memorial Hospital, Kaohsiung, Taiwan.
Background: Extracorporeal membrane oxygenation (ECMO) is a critical treatment for severe cardiopulmonary failure. However, traditional ECMO decannulation methods, such as manual compression and surgical repair, are associated with significant complications. This study evaluates suture-mediated closure devices, specifically Perclose ProGlide, as a potentially favorable decannulation strategy.
View Article and Find Full Text PDFInt J Cardiol Cardiovasc Risk Prev
March 2025
Department of Internal Medicine, University of New Mexico Health Sciences Center, Albuquerque, NM, USA.
Background: Transcatheter aortic valve replacement (TAVR) has significantly advanced the treatment of severe aortic stenosis (AS), particularly in elderly patients who often have coexisting coronary artery disease (CAD). Chronic total occlusion (CTO), a severe form of CAD, may negatively impact outcomes in TAVR patients, though data are limited. This meta-analysis aims to evaluate the impact of CTO on TAVR outcomes.
View Article and Find Full Text PDFJ Spine Surg
December 2024
Department of Neurosurgery, Geisinger Neuroscience Institute, Danville, PA, USA.
Anterior lumbar interbody fusion (ALIF) is an anterior surgical approach for interbody fusion in the lumbar spine which affords the surgeon unfettered access to the disc space and allows for release of the anterior longitudinal ligament and insertion of a large, lordotic interbody graft. Despite the benefits associated with ALIF when compared with other lumbar interbody fusion techniques, the ALIF approach is associated with a number of unique complications, and certain patient-specific criteria (e.g.
View Article and Find Full Text PDFJ Spine Surg
December 2024
Spine Clinic, Elsan Jean Villar Private Hospital, Bordeaux, France.
Background: Diffuse idiopathic skeletal hyperostosis (DISH) is a systemic condition that might lead to dysphagia because of massive osteophytes that may be present at multiple levels. Confirming the symptomatic level to guide surgical management and avoid extensive surgery is important, however, there is no globally accepted consensus on the topic.
Case Description: We report the case of a 51-year-old man, with no specific past medical history, who has been complaining of a 3-months pain in the left side of the tongue base with sensation of a lump in the throat and dysphagia.
Ecancermedicalscience
November 2024
Medical Oncology, General Hospital, PO Box 5408, Yaoundé, Cameroon.
The relationship between cancer and thrombosis was initially highlighted in the 19th century. Vascular complications in oncology can be arterial or venous thrombosis, and incidental pulmonary embolism is a growing challenge. We aimed to describe the frequency and clinical characteristics of cancer patients with incidental venous thromboembolism (iVTE).