Objective(s): To describe a new and more simple surgical procedure for implanting the Vibrant Soundbridge (VSB) on the short process of the incus in children with ear atresia and atretic plate and present the audiometric results.

Methods: Since 2014, pre- and post-operative audiometric tests with tonal and vocal audiometryat the maximal follow up, respectively 33, 22 and 12 months were performed after VSB implantation at the ENT department, Necker Enfants Malades Hospital, Paris, France. 3 children aged 11, 9 and 15 years with conductive hearing loss due to high grade ear atresia and absent ear canal were implanted. The malformed and fused malleus-incus complex was found to be immobile in two of the subjects, who then underwent a superior tympanotomy to carefully remobilize the malleus. Stapes were mobile in all cases. The FMT clip was customized and crimped on the short process, stabilized by the metallic wire.

Results: At the maximum follow-up time, the bone conduction was unchanged, and the mean aided ACPTA was 21 dB, 29 dB and 30 dB, compared to 66 dB, 63 dB and 68 dB unaided, respectively. The word recognition score (WRS) at 65 dB SPL increased from 10%, 10% and 70% (unaided) respectively to 100% (aided).

Conclusions: The long process of the incus is usually very hypoplastic in patients with ear atresia with a fused malleus-incus complex lateral to the stapes. VSB coupling to the long process of the incus or the stapes may be challenging in small mastoids. Coupling to the short process of the incus is comparatively a simple procedure, since there is no need for posterior tympanotomy and has no proximity to the facial nerve. The results of this study are similar to other ear atresia cases in the literature with classic FMT placement on the stapes or long process of the incus. Outcomes should be confirmed in a larger number of patients.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ijporl.2016.12.038DOI Listing

Publication Analysis

Top Keywords

process incus
24
ear atresia
20
short process
16
long process
12
vsb implantation
8
incus children
8
children ear
8
fused malleus-incus
8
malleus-incus complex
8
process
7

Similar Publications

Objectives: Although various stapedotomy and stapedectomy techniques exist, anchoring the piston can be challenging. We present a novel surgical approach for treating congenital stapes malformations with an atypical facial nerve trajectory.

Methods: This is a case of a 7-year-old boy presenting with bilateral conductive hearing loss.

View Article and Find Full Text PDF

In this study, a method for determining the optimal location and orientation of an implantable piezoelectric accelerometer on the short process of the incus is presented. The accelerometer is intended to be used as a replacement for an external microphone to enable totally implantable auditory prostheses. The optimal orientation of the sensor and the best attachment point are determined based on two criteria-maximum pressure sensitivity sum and minimum loudness level sum.

View Article and Find Full Text PDF

Background: The "floating mass transducer" (FMT) of the Vibrant Soundbridge was initially attached to the long process of the incus (LP). In patients missing an incus, coupling to the stapes head using the Vibroplasty CliP-coupler is used. Its main limitation is the height, leading to potential contact with the tympanic membrane.

View Article and Find Full Text PDF

Purpose: This study aims to analyze the imaging features of isolated congenital middle ear malformation (CMEM) on high-resolution computed tomography (HRCT).

Methods: We retrospectively collected patients with surgically confirmed diagnosis of isolated CMEM in our hospital between January 2018 and June 2023. All patients underwent HRCT before surgery.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!