Objective: Transcatheter aortic valve replacement is established therapy for high-risk and inoperable patients with severe aortic stenosis, but questions remain regarding long-term durability. Valve design influences durability. Increased leaflet stresses in surgical bioprostheses have been correlated with degeneration; however, transcatheter valve leaflet stresses are unknown. From 2007 to 2014, a majority of US patients received first-generation balloon-expandable transcatheter valves. Our goal was to determine stent and leaflet stresses in this valve design using finite element analyses.

Methods: A 26-mm Sapien Transcatheter Heart Valve (Edwards Lifesciences, Inc, Irvine, Calif) underwent high-resolution microcomputed tomography scanning to develop precise 3-dimensional geometry of the leaflets, the stent, and the polyethylene terephthalate elements. The stent was modeled using 3-dimensional elements and the leaflets were modeled using shell elements. Stent material properties were based on stainless steel, whereas those for leaflets were obtained from surgical bioprostheses. Noncylindrical Sapien valve geometry was also simulated. Pressure loading to 80 mm Hg and 120 mm Hg was performed using ABAQUS finite element software (Dassault Systèmes, Waltham, Mass).

Results: At 80 mm Hg, maximum principal stresses on Sapien leaflets were 1.31 megaspascals (MPa). Peak leaflet stress was observed at commissural tips where leaflets connected to the stent. Maximum principal stresses for the stent were 188.91 MPa and located at stent tips where leaflet commissures were attached. Noncylindrical geometry increased peak principal leaflet stresses by 16%.

Conclusions: Using exact geometry from high-resolution scans, the 26-mm Sapien Transcatheter Heart Valve showed that peak stresses for both stent and leaflets were present at commissural tips where leaflets were attached. These regions would be prone to leaflet degeneration. Understanding stresses in first-generation transcatheter valves allows comparison to future designs for relative durability.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5837867PMC
http://dx.doi.org/10.1016/j.jtcvs.2016.12.016DOI Listing

Publication Analysis

Top Keywords

leaflet stresses
20
stent
9
stresses
9
stent leaflet
8
first-generation balloon-expandable
8
balloon-expandable transcatheter
8
transcatheter aortic
8
valve
8
aortic valve
8
valve design
8

Similar Publications

Background And Objective: Deep vein thrombosis (DVT) of the lower limbs is a critical global vascular disease. Accurately assessing and predicting the efficacy of DVT treatment remains a significant challenge due to a lack of understanding of the mechanisms by which the level of patient-specific embolization and the rate of drug injection affect thrombolytic therapy.

Methods: In this study, we used the computed tomographic venography (CTV) clinical method to obtain patient-specific parameters, and the flow-solid interaction (FSI) method combined with biochemical response modeling of thrombolysis to analyze patient-specific hemodynamic and biomechanical characteristics and to quantitatively assess the effects of three vessel embolism levels (VEL) versus two drug injection rates (DIR) on bifurcated femoral venous thrombolytic therapy.

View Article and Find Full Text PDF

Lysophospholipids (LPLs) and host defense peptides (HDPs) are naturally occurring membrane-active agents that disrupt key membrane properties, including the hydrocarbon thickness, intrinsic curvature, and molecular packing. Although the membrane activity of these agents has been widely examined separately, their combined effects are largely unexplored. Here, we use experimental and computational tools to investigate how lysophosphatidylcholine (LPC) and lysophosphatidylethanolamine (LPE), an LPL of lower positive spontaneous curvature, influence the membrane activity of piscidin 1 (P1), an α-helical HDP from fish.

View Article and Find Full Text PDF

Objectives: The common surgical treatment in patients with obstructive hypertrophic cardiomyopathy (HCM) is septal myectomy. This involves resection of a segment of the myocardial septum and can be performed with and without concomitant anterior mitral valve leaflet extension (AMVLE). While both approaches have satisfying clinical outcomes, there is a lack of data regarding the added value of concomitant AMVLE.

View Article and Find Full Text PDF

Budding of Asymmetric Lipid Bilayers: Effects of Cholesterol, Anionic Lipid, and Electric Field.

J Phys Chem B

December 2024

Center for Statistical and Theoretical Condensed Matter Physics & Department of Physics, Zhejiang Normal University, Jinhua 321004, China.

Membrane budding is vital for various cellular processes such as synaptic activity regulation, vesicle transport and release, and endocytosis/exocytosis. Although protein-mediated membrane budding has been extensively investigated, the effects of the lipid asymmetry of the two leaflets and the asymmetrically electrical environments of the cellular membrane on membrane budding remain elusive. In this work, using coarse-grained molecular dynamics simulations, we systematically investigate the impacts of lipid bilayer asymmetry and external electric fields mimicking the asymmetric membrane potential on the membrane budding.

View Article and Find Full Text PDF

In several physiopathological processes, phosphatidylserine (PS), normally sequestered to the inner leaflet of the plasma membrane, becomes exposed to the cell surface. In erythrocytes (RBC), PS externalization is a crucial event for the removal of aged/damaged cells but can also be associated with increased prothrombotic activity. Structurally related olive oil antioxidants, including hydroxytyrosol (HT), are able to significantly reduce the percentage of PS-exposing RBC, when cells are exposed to toxic compounds such as the heavy metal mercury (Hg).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!