Purpose: To assess the reliability and validity of a hand motion sensor, Leap Motion Controller (LMC), in the 15-s hand grip-and-release test, as compared against human inspection of an external digital camera recording.

Methods: Fifty healthy participants were asked to fully grip-and-release their dominant hand as rapidly as possible for two trials with a 10-min rest in-between, while wearing a non-metal wrist splint. Each test lasted for 15 s, and a digital camera was used to film the anterolateral side of the hand on the first test. Three assessors counted the frequency of grip-and-release (G-R) cycles independently and in a blinded fashion. The average mean of the three was compared with that measured by LMC using the Bland-Altman method. Test-retest reliability was examined by comparing the two 15-s tests.

Results: The mean number of G-R cycles recorded was: 47.8 ± 6.4 (test 1, video observer); 47.7 ± 6.5 (test 1, LMC); and 50.2 ± 6.5 (test 2, LMC). Bland-Altman indicated good agreement, with a low bias (0.15 cycles) and narrow limits of agreement. The ICC showed high inter-rater agreement and the coefficient of repeatability for the number of cycles was ±5.393, with a mean bias of 3.63.

Conclusions: LMC appears to be valid and reliable in the 15-s grip-and-release test. This serves as a first step towards the development of an objective myelopathy assessment device and platform for the assessment of neuromotor hand function in general. Further assessment in a clinical setting and to gauge healthy benchmark values is warranted.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00586-017-4948-3DOI Listing

Publication Analysis

Top Keywords

grip-and-release test
8
digital camera
8
g-r cycles
8
lmc bland-altman
8
test lmc
8
test
7
hand
5
lmc
5
development assessment
4
assessment tool
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!