The biocompatibility and fixation of a new silicone intraocular lens was evaluated in the cat eye. Following extracapsular lens extraction, 14 cats were implanted with a silicone lens (SLM 2/UV type) with polypropylene modified J loops in one eye and a poly(methyl methacrylate) (PMMA) lens (Perspex CQ) of a similar design in the fellow eye. Half the lenses were placed in the ciliary sulcus and half in the capsular bag. The eyes were evaluated for up to one year. Neither lens material showed any signs of toxicity clinically or histopathologically. Both lenses achieved stable fixation in the capsular bag; however, some inflammatory reaction and lens dislocation were noted with sulcus placement of both lens types. The amount of proteinaceous and cellular debris on the explanted silicone lenses was significantly less than that on the PMMA lenses as assessed with methylene blue staining. All retrieved lenses were optically clear and no significant change in the lens surface morphology, clarity and/or optical properties was observed.