A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@gmail.com&api_key=61f08fa0b96a73de8c900d749fcb997acc09&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 176

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016

File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 316
Function: require_once

Comparison of loop-mediated isothermal amplification assay and smear microscopy with culture for the diagnostic accuracy of tuberculosis. | LitMetric

Comparison of loop-mediated isothermal amplification assay and smear microscopy with culture for the diagnostic accuracy of tuberculosis.

BMC Infect Dis

Department of Medical Microbiology, School of Biomedical and Laboratory sciences, CMHS, University of Gondar, P.O. box 196, Gondar, Ethiopia.

Published: January 2017

Background: Tuberculosis (TB) caused by Mycobacterium tuberculosis is one of the leading causes of death from infectious diseases worldwide. Sputum smear microscopy remains the most widely available pulmonary TB diagnostic tool particularly in resource limited settings. A highly sensitive diagnostic with minimal infrastructure, cost and training is required. Hence, we assessed the diagnostic performance of Loop-mediated isothermal amplification (LAMP) assay in detecting M.tuberculosis infection in sputum sample compared to LED fluorescent smear microscopy and culture.

Method: A cross-sectional study was conducted at the University of Gondar Hospital from June 01, 2015 to August 30, 2015. Pulmonary TB diagnosis using sputum LED fluorescence smear microscopy, TB-LAMP assay and culture were done. A descriptive analysis was used to determine demographic characteristics of the study participants. Analysis of sensitivity and specificity for smear microscopy and TB-LAMP compared with culture as a reference test was performed. Cohen's kappa was calculated as a measure of agreement between the tests.

Results: A total of 78 pulmonary presumptive TB patients sputum sample were analyzed. The overall sensitivity and specificity of LAMP were 75 and 98%, respectively. Among smear negative sputum samples, 33.3% sensitivity and 100% specificity of LAMP were observed. Smear microscopy showed 78.6% sensitivity and 98% specificity. LAMP and smear in series had sensitivity of 67.8% and specificity of 100%. LAMP and smear in parallel had sensitivity of 85.7% and specificity of 96%. The agreement between LAMP and fluorescent smear microscopy tests was very good (κ = 0.83, P-value ≤0.0001).

Conclusions: TB-LAMP showed similar specificity but a slightly lower sensitivity with LED fluorescence microscopy. The specificity of LAMP and smear microscopy in series was high. The sensitivity of LAMP was insufficient for smear negative sputum samples.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5240421PMC
http://dx.doi.org/10.1186/s12879-016-2140-8DOI Listing

Publication Analysis

Top Keywords

smear microscopy
32
specificity lamp
16
smear
12
lamp smear
12
microscopy
9
loop-mediated isothermal
8
isothermal amplification
8
lamp
8
sputum sample
8
fluorescent smear
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!