A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 176

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML

File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 316
Function: require_once

Repeatability and accuracy testing of a weight distribution platform and comparison to a pressure sensitive walkway to assess static weight distribution. | LitMetric

Repeatability and accuracy testing of a weight distribution platform and comparison to a pressure sensitive walkway to assess static weight distribution.

Vet Comp Orthop Traumatol

Prof. Simon C. Roe, North Carolina State University, College of Veterinary Medicine, Department of Clinical Sciences, 1052 William Moore Drive, Raleigh, NC 27607, United States, Phone: +1 919 513 6334, Fax: +1 919 513 6336, E-mail:

Published: March 2017

Objective: To evaluate the accuracy and repeatability of measurements collected using a weight distribution platform and a pressure sensitive walkway using an inanimate object with known weight distribution.

Methods: A custom-built jig with a range of weights was applied in a random order. Measurements were collected on both devices and compared to each other and to the known weight distribution.

Results: Weight distribution platform and pressure sensitive walkway measurements were highly correlated to each other (Pearson's correlation coefficient R = 0.98) and to actual weights (R = 0.99 for the weight distribution platform; 0.98 for the pressure sensitive walkway). Repeatability from day to day for both devices was greater than 0.99. For the weight distribution platform, the 95% confidence interval was ± 2.5% from the true percentage and ± 3.3% for the pressure sensitive walkway. The coefficient of variation (COV) was highest for both devices at the lightest weights (weight distribution platform 11.28%, pressure sensitive walkway 16.91%) and lowest with the heaviest weights (weight distribution platform 3.71%, pressure sensitive walkway 5.86%).

Conclusion: Both the weight distribution platform and the pressure sensitive walkway provided accurate and consistent measures of weight distribution with no significant difference between devices. The rounded standard error was three percent for the weight distribution platform, and four percent for the pressure sensitive walkway. The higher variability when measuring the smallest weight suggests less accuracy at lower weights with both devices.

Clinical Significance: The weight distribution platform is a repeatable and accessible device to measure static weight distribution, and if proven the same in a clinical setting, it will be a valuable addition to current objective measures of limb use.

Download full-text PDF

Source
http://dx.doi.org/10.3415/VCOT-16-09-0128DOI Listing

Publication Analysis

Top Keywords

weight distribution
52
distribution platform
40
pressure sensitive
36
sensitive walkway
36
weight
16
distribution
13
platform pressure
12
platform
10
pressure
9
sensitive
9

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!