Purpose: To determine whether lead-containing shields have lead dust on the external surface.
Methods: Institutional review board approval was obtained for this descriptive study of a convenience sample of 172 shields. Each shield was tested for external lead dust via a qualitative rapid on-site test and a laboratory-based quantitative dust wipe analysis, flame atomic absorption spectrometry (FAAS). The χ test was used to test the association with age, type of shield, lead sheet thickness, storage method, and visual and radiographic appearance.
Results: Sixty-three percent (95% confidence interval [CI]: 56%-70%) of the shields had detectable surface lead by FAAS and 50% (95% CI: 43%-57%) by the qualitative method. Lead dust by FAAS ranged from undetectable to 998 μg/ft. The quantitative detection of lead was significantly associated with the following: (1) visual appearance of the shield (1 = best, 3 = worst): 88% of shields that scored 3 had detectable dust lead; (2) type of shield: a greater proportion of the pediatric patient, full-body, and thyroid shields were positive than vests and skirts; (3) use of a hanger for storage: 27% of shields on a hanger were positive versus 67% not on hangers. Radiographic determination of shield intactness, thickness of interior lead sheets, and age of shield were unrelated to presence of surface dust lead.
Conclusions: Sixty-three percent of shields had detectable surface lead that was associated with visual appearance, type of shield, and storage method. Lead-containing shields are a newly identified, potentially widespread source of lead exposure in the health industry.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.jacr.2016.10.024 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!