Objective: The present study aimed to evaluate the accuracy of the automated oscillometric upper-arm blood pressure (BP) monitor BPUMP BF1112 for home BP monitoring according to the International Protocol of the European Society of Hypertension revision 2010 (ESH-IP2010).
Participants And Methods: Systolic and diastolic BPs were sequentially measured in 33 adult Chinese (13 women, mean age 46.7 years) using a mercury sphygmomanometer (two observers) and the BF1112 device (one supervisor). A total of 99 pairs of comparisons were obtained from 33 participants for judgments in two parts with three grading phases.
Results: The BPUMP BF1112 device achieved the targets in part 1 of the validation study. The number of absolute differences between device and observers within 5, 10, and 15 mmHg was 85/99, 96/99, and 97/99, respectively, for systolic BP, and 83/99, 97/99, and 99/99, respectively, for diastolic BP. The device also fulfilled the criteria in part 2 of the validation study. A total of 31 and 30 participants for systolic and diastolic BP, respectively, had at least two of the three device-observer differences within 5 mmHg (required≥24mmHg). No participant for systolic or diastolic BP had all the three device-observer comparisons greater than 5 mmHg.
Conclusion: The BPUMP BP monitor BF1112 has passed the requirements of the ESH-IP2010, and hence can be recommended for home use in adults.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1097/MBP.0000000000000236 | DOI Listing |
Blood Press Monit
April 2017
Shanghai Key Laboratory of Hypertension, Department of Hypertension, Centre for Epidemiological Studies and Clinical Trials, The Shanghai Institute of Hypertension, Ruijin Hospital, Shanghai Jiaotong University School of Medicine, Shanghai, China.
Objective: The present study aimed to evaluate the accuracy of the automated oscillometric upper-arm blood pressure (BP) monitor BPUMP BF1112 for home BP monitoring according to the International Protocol of the European Society of Hypertension revision 2010 (ESH-IP2010).
Participants And Methods: Systolic and diastolic BPs were sequentially measured in 33 adult Chinese (13 women, mean age 46.7 years) using a mercury sphygmomanometer (two observers) and the BF1112 device (one supervisor).
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!