Background: Segmented cine imaging with a steady-state free-precession sequence (Cine-SSFP) is currently the gold standard technique for measuring ventricular volumes and mass, but due to multi breath-hold (BH) requirements, it is prone to misalignment of consecutive slices, time consuming and dependent on respiratory capacity. Real-time cine avoids those limitations, but poor spatial and temporal resolution of conventional sequences has prevented its routine application. We sought to examine the accuracy and feasibility of a newly developed real-time sequence with aggressive under-sampling of k-space using sparse sampling and iterative reconstruction (Cine-RT).

Methods: Stacks of short-axis cines were acquired covering both ventricles in a 1.5T system using gold standard Cine-SSFP and Cine-RT. Acquisition parameters for Cine-SSFP were: acquisition matrix of 224×196, temporal resolution of 39ms, retrospective gating, with an average of 8 heartbeats per slice and 1-2 slices/BH. For Cine-RT: acquisition matrix of 224×196, sparse sampling net acceleration factor of 11.3, temporal resolution of 41ms, prospective gating, real-time acquisition of 1 heart-beat/slice and all slices in one BH. LV contours were drawn at end diastole and systole to derive LV volumes and mass.

Results: Forty-one consecutive patients (15 male; 41±17years) in sinus rhythm were successfully included. All images from Cine-SSFP and Cine-RT were considered to have excellent quality. Cine-RT-derived LV volumes and mass were slightly underestimated but strongly correlated with gold standard Cine-SSFP. Inter- and intra-observer analysis presented similar results between both sequences.

Conclusions: Cine-RT featuring sparse sampling and iterative reconstruction can achieve spatial and temporal resolution equivalent to Cine-SSFP, providing excellent image quality, with similar precision measurements and highly correlated and only slightly underestimated volume and mass values.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.mri.2017.01.002DOI Listing

Publication Analysis

Top Keywords

sparse sampling
16
temporal resolution
16
sampling iterative
12
iterative reconstruction
12
gold standard
12
cine imaging
8
volumes mass
8
spatial temporal
8
standard cine-ssfp
8
cine-ssfp cine-rt
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!