Fluorophore saturation is the key factor limiting the speed and excitation range of fluorescence lifetime imaging microscopy (FLIM). For example, fluorophore saturation causes incorrect lifetime measurements when using conventional frequency-domain FLIM at high excitation powers. In this Letter, we present an analytical theoretical description of this error and present a method for compensating for this error in order to extract correct lifetime measurements in the limit of fluorophore saturation. We perform a series of simulations and experiments to validate our methods. The simulations and experiments show a 13.2× and a 2.6× increase in excitation range, respectively. The presented method is based on algorithms that can be easily applied to existing FLIM setups.

Download full-text PDF

Source
http://dx.doi.org/10.1364/OL.42.000155DOI Listing

Publication Analysis

Top Keywords

fluorophore saturation
12
fluorescence lifetime
8
lifetime imaging
8
imaging microscopy
8
excitation range
8
lifetime measurements
8
simulations experiments
8
saturation-compensated measurements
4
measurements fluorescence
4
lifetime
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?