A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 176

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML

File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 316
Function: require_once

A novel adjustable dynamic plate for treatment of long bone fractures: An in vitro biomechanical study. | LitMetric

A novel adjustable dynamic plate for treatment of long bone fractures: An in vitro biomechanical study.

J Clin Orthop Trauma

Dokuz Eylul University, Faculty of Medicine, Department of Orthopaedics and Traumatology, 35340 Balcova, Izmir, Turkey.

Published: August 2016

Introduction: Locking compression plate (LCP) system was designed to provide bone stability and to enhance bone healing. However, implant failure, nonunion and instability are still frequently encountered complications. The purpose of this study was to assess and compare the biomechanical characteristics of a novel adjustable dynamic plate (ADP) with the commonly used LCP.

Materials And Methods: Twelve 4th generation composite artificial femoral bones were used. Transverse fracture was created in all bones, 6 femurs were fixated using the novel ADP, whereas the other 6 femurs were fixated using the traditional LCP. All samples had undergone a non-destructive repetitive different forces (axial compression, bending and torsion), to evaluate the biomechanical differences between the two plating systems.

Results: Under axial load the mean stiffness value was 439.0 N/mm for the ADP and 158.9 N/mm for the LCP, ADP showed a statistically significant stiffness value than LCP with a value of 0.004. There was no significant difference in flexion/extension bending strain values between ADP and LCP. However LCP provided significantly stiffer fixation in medial and lateral bending tests than ADP ( = 0.037) and ( = 0.016) respectively. But no significant difference was detected between the two plating system in the applied torsional stress.

Conclusion: These results do not show any significant biomechanical difference in the applied torsional and bending stresses between LCP and ADP. However the remarkably increased persistent compression effect of the ADP created a considerable stress on fracture edges which may accelerate bone healing.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5197444PMC
http://dx.doi.org/10.1016/j.jcot.2016.08.006DOI Listing

Publication Analysis

Top Keywords

novel adjustable
8
adjustable dynamic
8
dynamic plate
8
bone healing
8
adp
8
femurs fixated
8
lcp adp
8
applied torsional
8
lcp
7
plate treatment
4