The stiff Lunderquist guidewire may occasionally be needed during percutaneous insertion of a nephrostomy catheter, especially in patients with a large amount of subcutaneous fat or with a hard renal capsule. This wire may fracture at the junction of its stiff portion with the flexible tip. We have reported such an occurrence, in which an operative procedure was avoided by successful percutaneous retrieval of the broken wire tip, using the loop-snare technique.

Download full-text PDF

Source
http://dx.doi.org/10.1097/00007611-198910000-00017DOI Listing

Publication Analysis

Top Keywords

percutaneous retrieval
8
retrieval broken
8
lunderquist guidewire
8
broken lunderquist
4
guidewire pelvicalyceal
4
pelvicalyceal system
4
system stiff
4
stiff lunderquist
4
guidewire occasionally
4
occasionally needed
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!