Background: The chimney technique has been successfully used to treat juxtarenal aortic aneurysms. The two main issues with this technique are gutter formation and chimney graft (CG) compression, which induce a risk for type Ia endoleaks and stent thrombosis, respectively. In this benchtop study, the geometry and renal artery flow of chimney endovascular aneurysm repair configurations were compared with chimney configurations with endovascular aneurysm sealing (ch-EVAS).

Methods: Seven flow phantoms were constructed, including one control and six chimney endovascular aneurysm repairs (Endurant [Medtronic Inc, Minneapolis, Minn] and AFX [Endologix Inc, Irvine, Calif]) or ch-EVAS (Nellix, Endologix) configurations, combined with either balloon-expandable or self-expanding CGs with an intended higher positioning of the right CG in comparison to the left CG. Geometric analysis was based on measurements at three-dimensional computed tomography angiography and included gutter volume and CG compression, quantified by the ratio between maximal and minimal diameter (D-ratio). In addition, renal artery flow was studied in a physiologic flow model and compared with the control.

Results: The average gutter volume was 343.5 ± 142.0 mm, with the lowest gutter volume in the EVAS-Viabahn (W. L. Gore & Associates, Flagstaff, Ariz) combination (102.6 mm) and the largest in the AFX-Advanta V12 (Atrium Medical Corporation, Hudson, NH) configuration (559.6 mm). The maximum D-ratio was larger in self-expanding CGs than in balloon-expandable CGs in all configurations (2.02 ± 0.34 vs 1.39 ± 0.13). The CG compression had minimal influence on renal volumetric flow (right, 390.7 ± 29.4 mL/min vs 455.1 mL/min; left, 423.9 ± 28.3 mL/min vs 410.0 mL/min in the control).

Conclusions: This study showed that gutter volume was lowest in ch-EVAS in combination with a Viabahn CG. CG compression was lower in configurations with the Advanta V12 than with Viabahn. Renal flow is unrestricted by CG compression.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jvs.2016.10.058DOI Listing

Publication Analysis

Top Keywords

endovascular aneurysm
20
gutter volume
16
chimney endovascular
12
gutter formation
8
renal flow
8
flow chimney
8
aneurysm repair
8
aneurysm sealing
8
renal artery
8
artery flow
8

Similar Publications

: To study the influence of diabetes mellitus (DM) and metformin treatment on aneurysm sac remodeling after endovascular aneurysm repair (EVAR). : A retrospective single-center cohort analysis was conducted on consecutive patients who underwent elective EVAR for an infrarenal abdominal aortic aneurysm (AAA) between January 2011 and December 2021. Differences between study groups were analyzed and Kaplan-Meier analysis were employed to describe overall and reintervention-free survival.

View Article and Find Full Text PDF

: The parallel stent graft endovascular aortic repair (PGEVAR) technique is an off-the-shelf option used for elective complex abdominal aortic aneurysm repair with acceptable outcome results, as reported so far. The PGEVAR technique, using chimney or periscope parallel grafts, can also be used for patients with ruptured complex abdominal aortic aneurysms. However, only few data about the mid- to long-term outcomes are available.

View Article and Find Full Text PDF

Few clinical studies have reported on physician-modified inner-branched endovascular repair (PMiBEVAR) for aortic arch aneurysm. Herein, we evaluate the outcomes of proximal landing zone 2 PMiBEVAR. This retrospective study analyzed data from six patients who underwent zone 2 PMiBEVAR for aortic arch aneurysms at a single center between October 2021 and June 2024.

View Article and Find Full Text PDF

Endovascular stent graft repair was developed to minimize the invasiveness of open surgery for thoracic and abdominal aortic diseases. This approach involves covering the diseased segment with a stented artificial graft. However, in thoracic endovascular aortic repair (TEVAR) for aortic arch diseases, special consideration is needed to preserve the aortic arch vessels.

View Article and Find Full Text PDF

Thoracoabdominal aortic aneurysms (TAAAs) are rare but serious conditions characterized by dilation of the aorta characterized by remodeling of the vessel wall, with changes in the elastin and collagen content. Individuals with Marfan syndrome have a genetic predisposition for elastic fiber fragmentation and elastin degradation and are prone to early aneurysm formation and progression. Our objective was to analyze the medial collagen characteristics through histological, polarized light microscopy, and electron microscopy methods across the thoracic and abdominal aorta in twenty-five patients undergoing open surgical repair, including nine with Marfan syndrome.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?