Introduction: In the assessment of hand and upper limb function, grip strength is of major importance. Use of the JAMAR dynamometer has been an accepted test of grip strength and routinely part of the physical examination.
Purpose Of The Study: The aim of this study was to determine by means of a median nerve block simulating a patient group if the measurement of different types of grip force can improve the quantification of impairment of the hand beyond the sole power grip measurements.
Methods: In 29 healthy volunteers, grip force measurement was done by the JAMAR dynamometer for power grip and by a pinch grip dynamometer for pincer grip, precision grip, pinch grip, and palmar abduction. At the power grip, the force of digit 2-5 was measured separately by a sensor glove. For each measurement, 3 trials were recorded subsequently.
Study Design: A within-subject research design was used in this prospective study. Each subject served as the control by preinjection measurements of grip and pinch strength. Subsequent measurements after median nerve block were used to examine within-subject change.
Results: In power grip, there was no significant reduction of maximum grip force with median nerve block compared with grip force without block. By measuring the finger forces by a sensor glove, a reduction of force at the individual fingers was found compared with the setting with no block, although significant only at the middle finger (P < .017). With median nerve block, average grip force decreased significantly 20% in pincer grip (P < .021), 29.5% in pinch grip (P < .002), 39.5% in precision grip (P < .001), and 70.7% in palmar abduction (P < .013).
Conclusions: These experiments suggest a minor role of the measurement of the power grip force in the diagnostics of distal median nerve block with the dynamometer, which could not be substantially improved by the digit forces and the force distribution of the individual fingers. The assessment of other grip forms, such as pincer grasp, precision grip, pinch grip, and especially palmar abduction, leads to a relevant improvement in the diagnostics of distal median nerve lesion.
Level Of Evidence: II.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.jht.2016.11.004 | DOI Listing |
Int Urol Nephrol
January 2025
Department of Urology, Faculty of Medicine, Cairo University, Cairo, Egypt.
Objective: To evaluate the outcomes and efficacy of robot-assisted radical prostatectomy (RARP) using the Versius robotic surgical system, aiming to provide comprehensive data on perioperative outcomes, postoperative recovery, and complications.
Patient And Methods: All cases of RARP using the CMR Versius platform performed at Cairo University Hospital over a two-year period were enrolled in this study. All patients had pathologically confirmed prostate cancer in both localized and locally advanced stages.
Int Orthop
January 2025
Physical Medicine and Rehabilitation Division, Department of Orthopaedic Surgery, Stanford University, Redwood City, CA, USA.
Purpose: This narrative review identifies and summarizes current evidence for diagnostic ultrasonographic evaluation of upper extremity dynamic compressive neuropathies affecting athletes.
Methods: Relevant literature was identified using the PubMed database and then summarized.
Results: The compressive neuropathies affecting athletes we identified included: neurogenic thoracic outlet syndrome, pectoralis minor syndrome, quadrilateral space syndrome, suprascapular nerve entrapment, proximal median nerve entrapment or bicipital aponeurosis/lacertus fibrosus (lacertus syndrome), radial tunnel syndrome, and cubital tunnel syndrome.
Background: Understanding the median nerve's position relative to surrounding anatomy is essential; however, there are many variations among individuals. This study assesses differences in median nerve position with or without palmaris longus (PL). We hypothesize that PL presence alters median nerve position, resulting in a greater distance to the skin volar surface, a decreased distance to the radius volar surface, and an increased distance to the flexor carpi radialis (FCR).
View Article and Find Full Text PDFHypertension
January 2025
Cardiology Department (P.B., X.L., V.T.T., M.A.B., A.V., E.Y., D.M.N., U.P., J.L., S.P.T., P.C.Q.), Westmead Hospital, Sydney, Australia.
Background: Transcatheter renal denervation (RDN) remains inconsistent despite developments in ablation technologies, due to the lack of an intraprocedural physiological end point.
Objective: To identify whether aorticorenal ganglion (ARG) guided RDN using microwave (MW) catheter leads to more consistent denervation outcomes compared with empirical MW ablation.
Methods: Pigs underwent sham procedure (n=8) or bilateral RDN using an in-house built open-irrigated MW catheter.
J Neurol Surg B Skull Base
February 2025
Department of Neurosurgery, UC San Diego Medical Center, San Diego, California, United States.
Many patients with neurofibromatosis type 2 (NF2) suffer from sensorineural hearing loss, and associated cochlear nerve compromise in NF2 patients makes auditory brainstem implant (ABI) an attractive treatment option. The long-term outcomes and benefits of the device are still being explored. A retrospective review was conducted for 11 ABI recipients at a single-institution tertiary center between November 2017 and August 2022.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!