Objective: We investigate the potential of a prototype multimodality arthroscope, combining ultrasound, optical coherence tomography (OCT) and arthroscopic indentation device, for assessing cartilage lesions, and compare the reliability of this approach with conventional arthroscopic scoring ex vivo.

Design: Areas of interest (AIs, N = 43) were selected from equine fetlock joints (N = 5). Blind-coded AIs were independently scored by two equine surgeons employing International Cartilage Repair Society (ICRS) scoring system via conventional arthroscope and multimodality arthroscope, in which high-frequency ultrasound and OCT catheters were attached to an arthroscopic indentation device. In addition, cartilage stiffness was measured with the indentation device, and lesions in OCT images scored using custom-made automated software. Measurements and scorings were performed twice in two separate rounds. Finally, the scores were compared to histological ICRS scores.

Results: OCT and arthroscopic examinations showed the highest average agreements (55.2%) between the scoring by surgeons and histology scores, whereas ultrasound had the lowest (50.6%). Average intraobserver agreements of surgeons and interobserver agreements between rounds were, respectively, for conventional arthroscope (68.6%, 69.8%), ultrasound (68.6%, 68.6%), OCT (65.1%, 61.7%) and automated software (65.1%, 59.3%).

Conclusions: OCT imaging supplemented with the automated software provided the most reliable lesion scoring. However, limited penetration depth of light limits the clinical potential of OCT in assessing human cartilage thickness; thus, the combination of OCT and ultrasound could be optimal for reliable diagnostics. Present findings suggest imaging and quantitatively analyzing the entire articular surface to eliminate surgeon-related variation in the selection of the most severe lesion to be scored.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.joca.2016.12.007DOI Listing

Publication Analysis

Top Keywords

indentation device
12
automated software
12
equine fetlock
8
multimodality arthroscope
8
oct
8
oct arthroscopic
8
arthroscopic indentation
8
conventional arthroscope
8
ultrasound
5
multimodality scoring
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!