Introduction: Proper implantation of a hinged external elbow fixator (HEEF) is demanding since it requires precise alignment between the flexion-extension's and HEEF's axis. In order to optimize this alignment, we have developed a 3D-printed aiming device. The primary goal of the study was to compare the aiming device-based technique with the conventional pin technique. The secondary goal was to determine whether it is possible to share the aiming device with the surgical community.
Materials And Methods: A HEEF was implanted in cadavers with either the aiming device (n = 6) or the conventional pin technique (n = 6). For both techniques the duration of the procedure, the radiation exposure as well as the offset and angular divergence between the HEEF's and flexion-extension's axis were compared. To achieve the secondary goal, two surgeons used aiming devices 3D-printed from files sent by email in order to implant HEEF on cadaveric specimens (n = 6).
Results: Duration of the procedure was not significantly different between both techniques. However, the aiming device allowed for reduction of the number of image intensifier shots (p = 0.005), angular divergence (p = 0.02) and offset between both axes (p = 0.05). The aiming devices have been delivered less than 15 days after ordering, and they have allowed proper implantation of six HEEF.
Conclusion: The 3D-printed aiming device allowed less irradiant and more accurate implantation of HEEF. It is possible to share it with other surgeons.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1007/s00590-016-1889-1 | DOI Listing |
Sci Rep
December 2024
Department of Medical Device Development, Seoul National University College of Medicine, Seoul, Republic of Korea.
Vertebral collapse (VC) following osteoporotic vertebral compression fracture (OVCF) often requires aggressive treatment, necessitating an accurate prediction for early intervention. This study aimed to develop a predictive model leveraging deep neural networks to predict VC progression after OVCF using magnetic resonance imaging (MRI) and clinical data. Among 245 enrolled patients with acute OVCF, data from 200 patients were used for the development dataset, and data from 45 patients were used for the test dataset.
View Article and Find Full Text PDFSci Rep
December 2024
Department of Pediatrics and Child Health Nursing, College of Medicine and Health Sciences, Bahir Dar, Ethiopia.
Introduction: Insecticide-treated bed nets are often used as a physical barrier to prevent infection of malaria. In Sub-Saharan Africa, one of the most important ways of reducing the malaria burden is the utilization of insecticide-treated bed nets. However, there is no sufficient information on the utilization of insecticide-treated bed nets and their associated factors in Ethiopia.
View Article and Find Full Text PDFBAY 2413555 is a novel selective and reversible positive allosteric modulator of the type 2 muscarinic acetylcholine (M2) receptor, aimed at enhancing parasympathetic signaling and restoring cardiac autonomic balance for the treatment of heart failure (HF). This study tested the safety, tolerability and pharmacokinetics of this novel therapeutic option. REMOTE-HF was a multicenter, double-blind, randomized, placebo-controlled, phase Ib dose-titration study with two active arms.
View Article and Find Full Text PDFJMIR Rehabil Assist Technol
December 2024
Centre de recherche interdisciplinaire en réadaptation du Montréal métropolitain (CRIR) - Institut universitaire sur la réadaptation en déficience physique de Montréal (IURDPM) du Centre intégré universitaire de santé et de services sociaux du Centre-Sud-de-l'Île-de-Montréal (CCSMTL), Université de Montréal, Institut de Réadaptation Gingras Lindsay de Montréal, 6300 avenue de Darlington, Montréal, QC, H3S 2J4, Canada, 1 514-343-6111.
Background: Stationary bikes are used in numerous rehabilitation settings, with most offering limited functionalities and types of training. Smart technologies, such as artificial intelligence and robotics, bring new possibilities to achieve rehabilitation goals. However, it is important that these technologies meet the needs of users in order to improve their adoption in current practice.
View Article and Find Full Text PDFCatheter Cardiovasc Interv
December 2024
Department of Cardiology, Angiology and Intensive Care Medicine, University Heart Center Lübeck, Lübeck, Germany.
Background: The new Amplatzer Steerable Delivery Sheath is a delivery system designed to improve ease-of-use and procedural results of left atrial appendage closure (LAAC). We aimed to compare procedural results after switching our LAAC program at a tertiary care center with the Amulet device to the Steerable Delivery Sheath, with a control group of LAAC employing the standard sheath.
Methods: The first n = 32 consecutively treated patients at our site using the Amulet device with the Steerable Delivery Sheath were included in this retrospective analysis.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!