Background: Five key factors enabling a good surgical grossing technique include a flat uniformly perpendicular specimen cutting face, appropriate immobilisation of the tissue specimen during grossing, good visualisation of the cutting tissue face, sharp cutting knives and the grossing knife action. TruSlice and TruSlice Digital are new innovative tools based on a guillotine configuration. The TruSlice has plastic inserts whilst the TruSlice Digital has an electronic micrometre attached: both features enable these dissection factors to be controlled. The devices were assessed in five hospitals in the UK.

Material And Methods: A total of 267 fixed tissue samples from 23 tissue types were analysed, principally the breast (n = 32) skin (30), rectum (28), colon (27) and cervix (17). Precision and accuracy were evaluated by measuring the defined thickness, and the consistency of achieving the defined thickness of tissue samples taken respectively. Both parameters were expressed as a total percentage of compliance for the cohort of samples accessed.

Results: Overall, the mean (standard deviation) score for precision was 81 (11) % whilst the accuracy score was 82 (11) % (both p < 0.05, chi-squared test), although this varied with type of tissue. Accuracy and precision were strongly correlated (r = 0.83, p < 0.001).

Conclusion: The TruSlice Digital devices offer an assured precision and accuracy performance which is reproducible across an assortment of tissue types. The use of a micrometre to set tissue slice thickness is innovative and should comply with laboratory accreditation requirements, alleviating concerns of how to tackle issues such as the 'measurement of uncertainty' at the grossing bench.

Download full-text PDF

Source
http://dx.doi.org/10.1080/09674845.2016.1233791DOI Listing

Publication Analysis

Top Keywords

truslice digital
12
precision accuracy
8
truslice truslice
8
tissue samples
8
defined thickness
8
truslice
6
tissue
5
multicentre study
4
study precision
4
accuracy truslice
4

Similar Publications

Background: Five key factors enabling a good surgical grossing technique include a flat uniformly perpendicular specimen cutting face, appropriate immobilisation of the tissue specimen during grossing, good visualisation of the cutting tissue face, sharp cutting knives and the grossing knife action. TruSlice and TruSlice Digital are new innovative tools based on a guillotine configuration. The TruSlice has plastic inserts whilst the TruSlice Digital has an electronic micrometre attached: both features enable these dissection factors to be controlled.

View Article and Find Full Text PDF

Histological dissection of human tissue has relied on conventional procedures, which have largely remained unchanged for decades. Practices to determine measurement parameters employed in these procedures have largely relied on the use of rulers and weighing scales. It is well documented in the scientific literature that both fixation and processing of tissue can significantly affect the viability of the of tissue sections both for tinctorial and immunocytochemical investigations.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!