High resolution spectrometer for extended x-ray absorption fine structure measurements in the 6 keV to 15 keV energy range.

Rev Sci Instrum

Artep Inc., 2922 Excelsior Springs Court, Ellicott City, Maryland 21042, USA.

Published: November 2016

A Cauchois transmission-crystal spectrometer has been developed with high crystal resolving power in the 6 keV-15 keV energy range and sufficient sensitivity to record single-shot spectra from the Lawrence Livermore National Laboratory (LLNL) Titan laser and other comparable or more energetic lasers. The spectrometer capabilities were tested by recording the W L transitions from a laboratory source and the extended x-ray absorption fine structure (EXAFS) spectrum through a Cu foil.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5488268PMC
http://dx.doi.org/10.1063/1.4959918DOI Listing