We have designed and installed a new Langmuir-probe (LP) array diagnostic to determine basic three-dimensional (3D) features of plasmas in TORPEX. The diagnostic consists of two identical LP arrays, placed on opposite sides of the apparatus, which provide comprehensive coverage of the poloidal cross section at the two different toroidal locations. Cross correlation studies of signals from the arrays provide a basic way to extract 3D information from the plasmas, as experiments show. Moreover, the remarkable signal-to-noise performance of the front-end electronics allows us to follow a different approach in which we combine information from all probes in both arrays to reconstruct elementary 3D plasma structures at each acquisition time step. Then, through data analysis, we track the structures as they evolve in time. The LP arrays include a linear-motion mechanism that can displace radially the probes located on the low field side for experiments that require fine-tuning of the probe locations, and for operational compatibility with the recently installed in-vessel toroidal conductor.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1063/1.4968024 | DOI Listing |
Rev Sci Instrum
March 2024
Department of Electrical Engineering and Electronics, University of Liverpool, Brownlow Hill, Liverpool L69 3GJ, United Kingdom.
By utilizing Field Programmable Gate Arrays in a configuration similar to that of the Mirror Langmuir Probe, it is possible to bias a single probe at three precise voltages in sequence. These voltages can be dynamically adjusted in real-time based on the measured plasma electron temperature to ensure the transition region is always sampled. The first results have been obtained by employing this method and have generated real-time outputs of electron temperature, ion saturation current, and floating potential on a low temperature pulsed-DC magnetron at 500 kHz.
View Article and Find Full Text PDFRev Sci Instrum
December 2023
Institute of Plasma Physics, Hefei Institutes of Physical Science, Chinese Academy of Sciences, Hefei 230031, China.
A compact helicon plasma source for the study of helicon plasma, especially for the study of blue core plasma, is designed and developed with permanent magnets (PMs). The structure of the PMs consists of two sets of ring array magnets with opposite magnetization. This structure can provide a higher magnetic field with fewer PMs, which is helpful for controlling the device's mass.
View Article and Find Full Text PDFRev Sci Instrum
October 2021
Massachusetts Institute of Technology, Plasma Science and Fusion Center, 77 Massachusetts Ave., Cambridge, Massachusetts 02139, USA.
For the first time, a digital Mirror Langmuir Probe (MLP) has successfully sampled plasma temperature, ion saturation current, and floating potential together on a single probe tip in real time in a radio-frequency driven helicon linear plasma device. This is accomplished by feedback control of the bias sweep to ensure a good fit to I-V characteristics with a high frequency, high power digital amplifier, and field-programmable gate array controller. Measurements taken by the MLP were validated by a low speed I-V characteristic manually collected during static plasma conditions.
View Article and Find Full Text PDFRev Sci Instrum
April 2021
Swiss Plasma Center (SPC), École Polytechnique Fédérale de Lausanne (EPFL), CH-1015 Lausanne, Switzerland.
A detailed description of the construction and commissioning of the fast-moving reciprocating divertor Langmuir probe array on the Tokamak à Configuration Variable (TCV) is presented. A horizontal array of 24 Langmuir probes (12 Mach probes) combined with a fast vertical movement provides unprecedented two-dimensional measurements of the plasma properties across the entire divertor outer leg volume and up to the X-point. The L-shape probe arm has to be very compact to minimize plasma perturbations and, at the same time, has to house 24 coaxial transmission lines, withstand time-averaged heat fluxes of up to ≈30 MW/m and accelerations of up to 8 g, and be resilient to violent plasma disruptions.
View Article and Find Full Text PDFRev Sci Instrum
May 2021
Department of Physics, Queen's University, Kingston, Ontario K7L 3N6, Canada.
The radial electric field in a field-reversed configuration (FRC) plasma plays an important role in the global stability and confinement properties. Herein, we developed a new Langmuir probe array named "Skewered probe" employed in measuring the radial potential profile in the collisional merging formation of an FRC in the FAT-CM (FRC Amplification via Translation - Collisional Merging) device. Because an FRC has a strong toroidal flow, the skewered probe consists of alternately skewered ring electrodes and ceramic beads on a thin stainless-steel tube to neutralize the effect of plasma flow.