Objectives: The aim of this study was to examine the impact of 3-dimensional (3D) computed tomographic (CT) guided procedural planning for left atrial appendage (LAA) occlusion on the early operator WATCHMAN learning curve.

Background: Traditional WATCHMAN implantation is dependent on 2-dimensional transesophageal echocardiographic (TEE) sizing and intraprocedural guidance.

Methods: LAA occlusion with the WATCHMAN device was performed in 53 patients. Pre-procedural case plans were generated from CT studies with recommended device size, catheter selection, and C-arm angle for deployment.

Results: All 53 patients underwent successful LAA occlusion with the WATCHMAN. Three-dimensional CT LAA maximal-width sizing was 2.7 ± 2.2 mm and 2.3 ± 3.0 mm larger than 2-dimensional and 3D TEE measurements, respectively (p ≤ 0.0001). By CT imaging, device selection was 100% accurate. There were 4 peri-WATCHMAN leaks (<4.5 mm) secondary to accessory LAA pedunculations. By 2-dimensional TEE maximal-width measurements alone, 62.3% (33 of 53) would have required larger devices. Using 3D TEE maximal-width measurements, 52.8% of cases (28 of 53) would have required larger devices. Three-dimensional TEE length would have inappropriately excluded 10 patients from WATCHMAN implantation. Compared with the average of 1.8 devices used per implantation attempt in PROTECT AF (WATCHMAN Left Atrial Appendage System for Embolic Protection in Patients With Atrial Fibrillation) (82% success rate), the present site averaged 1.245 devices per implantation attempt (100% success rate). There were no intraprocedural screen failures and no major adverse cardiac events.

Conclusions: Three-dimensional CT image case planning provides a comprehensive and customized patient-specific LAA assessment that appears to be accurate and may possibly facilitate reducing the early WATCHMAN implantation learning curve.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jcin.2016.07.038DOI Listing

Publication Analysis

Top Keywords

laa occlusion
12
3-dimensional computed
8
computed tomographic
8
watchman implantation
8
occlusion watchman
8
watchman
5
application 3-dimensional
4
tomographic image
4
image guidance
4
guidance watchman
4

Similar Publications

Background Stroke is a major cause of death and disability globally, with different stroke burdens in different regions. This paper reviews the epidemiology of stroke in Asia. Summary There is a wide range in age and sex-standardised stroke incidence, highest in China, lowest in Bhutan.

View Article and Find Full Text PDF

Preclinical Experience Using 4D Intracardiac Echocardiography to Guide Cardiac Electrophysiology Procedures.

J Cardiovasc Electrophysiol

December 2024

Division of Cardiovascular Medicine, Cardiac Electrophysiology Section, Hospital of the University of Pennsylvania, Philadelphia, Pennsylvania, USA.

Introduction: Intracardiac echocardiography (ICE) is an essential imaging modality for electrophysiology procedures, allowing intraprocedural monitoring, real-time catheter manipulation guidance, and visualization of complex anatomic structures. Four-dimentional (4D) ICE is the next stage in the evolution of the technology, permitting 360° rotation of the imaging plane, simultaneous multiplanar imaging, and volumetric acquisition, similar to transesophageal echocardiography (TEE). In this study, we report our experience with a novel 4D ICE catheter (NuVision, Biosense Webster) in structural electrophysiology procedures and difficult ventricular ablations in a swine preclinical model.

View Article and Find Full Text PDF

Background: The new Amplatzer Steerable Delivery Sheath is a delivery system designed to improve ease-of-use and procedural results of left atrial appendage closure (LAAC). We aimed to compare procedural results after switching our LAAC program at a tertiary care center with the Amulet device to the Steerable Delivery Sheath, with a control group of LAAC employing the standard sheath.

Methods: The first n = 32 consecutively treated patients at our site using the Amulet device with the Steerable Delivery Sheath were included in this retrospective analysis.

View Article and Find Full Text PDF

Background: The congenital absence of the left atrial appendage (LAA) is an extremely rare anatomical anomaly, with only 23 cases documented in medical literature. The LAA plays a critical role in thrombus formation, particularly in patients with atrial fibrillation (AF), thus impacting stroke prevention strategies and the management of anticoagulation.

Case Summary: We report a 48-year-old male with a 2-year history of hypertension and prior episodes of tachycardic palpitations, who presented with AF and chest pain.

View Article and Find Full Text PDF

Background: Collateral circulation plays a key role in acute ischemic stroke. We sought to determine the association between the arterial collateral status, estimated by the Hypoperfusion Intensity Ratio (HIR) on perfusion MRI, and stroke etiology in anterior circulation large vessel occlusion (LVO).

Methods: We retrospectively analyzed anterior circulation LVO acute stroke patients with a baseline perfusion MRI performed within 24 h from symptom onset.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!