Objective: In previous studies, multitrack trajectories in deep brain stimulation (DBS) have usually been approximated. Using a geometrically more accurate method, we compared the stereotactic accuracy of DBS with multitrack microelectrode recording and awake stimulation (function group) and MRI-guided DBS (MRI group).

Methods: One hundred and seventy-two leads used in DBS between April 2014 and January 2016 were evaluated for stereotactic errors. Targets were the subthalamic nucleus (STN, 139 leads) or globus pallidus interna (GPi, 33 leads). We geometrically calculated shifted-track targets by considering the three-dimensional stereotactic ring and arc rotations. Stereotactic errors were calculated using Euclidean distances perpendicular to trajectories. Motor outcomes according to the Unified Parkinson's Disease Rating Scale (UPDRS) part III, improvement percentages by stimulations, were analyzed in 24 patients with 1 year follow-ups.

Results: Functional evaluation tended to increase stereotactic errors in the STN function group (n = 129; 1.4 ± 0.7 mm) more than in the STN MRI group (n = 10; 1.0 ± 0.6 mm; P = 0.06). Leads with higher stereotactic errors (n = 65; 1.6 ± 0.7 mm; P < 0.001) than the center-track leads (n = 64; 1.2 ± 0.7 mm) were selectively track shifted. Track-shifted leads approached MRI targets in 86% (56/65 leads). Lower stereotactic errors tended to correlate with a better outcome (P = 0.095). Distances to MRI targets >2.5 mm tended to relate to a worse outcome (P = 0.087). Stereotactic errors were lower (n = 33; 0.9 ± 0.5 mm) in the GPi MRI group.

Conclusions: Multitrack DBS using intraoperative functional evaluation resulted in worse stereotactic accuracy than did MRI-guided DBS. However, track shifts in function-guided DBS can approach MRI targets effectively.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.wneu.2016.11.046DOI Listing

Publication Analysis

Top Keywords

stereotactic errors
24
deep brain
12
brain stimulation
12
mri targets
12
stereotactic
9
stereotactic accuracy
8
function group
8
mri-guided dbs
8
functional evaluation
8
group n =
8

Similar Publications

Article Synopsis
  • Dynamic conformal arc therapy (DCAT) and volumetric modulated arc therapy (VMAT) both provide similar quality for treating brain metastases, but this study focused on how MLC errors during treatment affect dose distribution.
  • Researchers created reference plans and simulated MLC errors using a head phantom, analyzing how these errors influenced dose delivery to the targets and surrounding brain tissue.
  • The findings revealed that while MLC errors affected both techniques' dose distribution similarly, the DCAT technique showed greater resilience, with less variation in target doses compared to VMAT.
View Article and Find Full Text PDF

An MRI-guided stereotactic neurosurgical robotic system for semi-enclosed head coils.

J Robot Surg

December 2024

National Engineering Research Center of Neuromodulation, School of Aerospace Engineering, Tsinghua University, Beijing, 100084, China.

Magnetic resonance imaging (MRI) offers high-quality soft tissue imaging without radiation exposure, which allows stereotactic techniques to significantly improve outcomes in cranial surgeries, particularly in deep brain stimulation (DBS) procedures. However, conventional stereotactic neurosurgeries often rely on mechanical stereotactic head frames and preoperative imaging, leading to suboptimal results due to the invisibility and the contact with patient's head, which may cause additional harm. This paper presents a frameless, MRI-guided stereotactic neurosurgical robotic system.

View Article and Find Full Text PDF

Intrafraction Patient Positional Uncertainty in Lung Stereotactic Ablative Radiotherapy with Abdominal Compression.

Pract Radiat Oncol

December 2024

Department of Radiation Oncology, Willis Knighton Cancer Center, 2600 Kings Highway, Shreveport, Louisiana, USA 71103 &, Department of Clinical Research, University of Jamestown, Fargo, ND, USA. Electronic address:

Purpose: Motion management presents a significant challenge in thoracic stereotactic ablative radiotherapy (SABR). Currently, a 5.0 mm standard planning target volume (PTV) margin is widely used to ensure adequate dose to the tumor.

View Article and Find Full Text PDF

Primary Central Nervous System Lymphoma (PCNSL) Following Thyroid Cancer Surgery: A Case Report of Misdiagnosed Brain Metastasis and Literature Review.

Curr Oncol

November 2024

Department of Neurosurgery, Center for Malignant Brain Tumors, National Glioma MDT Alliance, Peking Union Medical College Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing 100730, China.

Unlabelled: This article reports a rare case of primary central nervous system lymphoma (PCNSL) found in a patient with thyroid cancer after surgery.

Methods: The patient was initially misdiagnosed with brain metastases, and the diagnosis of PCNSL was later confirmed by pathology.

Results: The analysis of this case and review of the relevant literature explores the possible mechanisms of the coexistence of thyroid cancer and PCNSL, as well as their diagnostic, differential diagnostic, and therapeutic challenges.

View Article and Find Full Text PDF

Introduction: Treatment of neoplasic lung nodules with ground glass opacities (GGO) faces two primary challenges. First, the standard practice of treating GGOs as solid nodules, which effectively controls the tumor locally, but might increase associated toxicities. The second is the potential for dose calculation errors related to increased heterogeneity.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!