Background: Electrical heterogeneity (EH) during cardiac resynchronization therapy may vary with different left ventricular (LV) pacing sites.

Objective: The purpose of this study was to evaluate the relationship between such changes and acute hemodynamic response (AHR).

Methods: Two EH metrics-standard deviation of activation times and mean left thorax activation times-were computed from isochronal maps based on 53-electrode body surface mapping during baseline AAI pacing and biventricular (BiV) pacing from different pacing sites in coronary veins in 40 cardiac resynchronization therapy-indicated patients. AHR at different sites was evaluated by invasive measurement of LV-dp/dt at baseline and BiV pacing, along with right ventricular (RV)-LV sensing delays and QRS duration.

Results: The site with the greatest combined reduction in standard deviation of activation times and left thorax activation times from baseline to BiV pacing was hemodynamically optimal (defined by AHR equal to, or within 5% of, the largest dp/dt response) in 35 of 40 patients (88%). Sites with the longest RV-LV and narrowest paced QRS were hemodynamically optimal in 26 of 40 patients (65%) and 28 of 40 patients (70%), respectively. EH metrics from isochronal maps had much better accuracy (sensitivity 90%, specificity 80%) for identifying hemodynamically responsive sites (∆LV dp/dt ≥10%) compared with RV-LV delay (69%, 85%) or paced QRS reduction (52%, 76%). Multivariate prediction model based on EH metrics showed significant correlation (R = 0.53, P <.001) between predicted and measured AHR.

Conclusion: Changes in EH from baseline to BiV pacing more accurately identified hemodynamically optimal sites than RV-LV delays or paced QRS shortening. Optimization of LV lead location by minimizing EH during BiV pacing, based on body surface mapping, may improve CRT response.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.hrthm.2016.11.017DOI Listing

Publication Analysis

Top Keywords

cardiac resynchronization
12
activation times
12
biv pacing
12
body surface
8
surface mapping
8
electrical heterogeneity
8
left ventricular
8
ventricular pacing
8
pacing sites
8
acute hemodynamic
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!