Intraocular pressure measurement after DSAEK by iCare, Goldmann applanation and dynamic contour tonometry: A comparative study.

J Fr Ophtalmol

Department of Ophthalmology, The Edith Wolfson Medical Center, Holon, Israel; Sackler School of Medicine, Tel-Aviv University, Tel Aviv, Israel.

Published: December 2016

Purpose: Corneal thickness inevitably increases following Descemet's stripping automated endothelial keratoplasty (DSAEK), owing to the addition of a donor graft. The current study compares different devices in assessing post-DSAEK intraocular pressure (IOP).

Methods: We compared IOP values measured by the Goldmann tonometry (GAT), iCare rebound tonometry (iCare) and Pascal dynamic contour tonometry (PDCT) in eyes following DSAEK. Agreement between measurements was calculated with correlation analysis and Bland-Altman plots. Effects of keratometry, central, thickness (CCT), endothelial cell density (ECD) and axial length on IOP measurements were assessed with Pearson's correlation.

Results: Twenty eyes of 20 patients (mean age 74.3±14.4, 14 females) post-DSAEK were included in this study. There was a high concordance between the IOP readings obtained by the three devices: a strong and significant correlation was found between GAT and PDCT (r=0.94, P<0.001) GAT and iCare (r=0.86, P<0.001) and iCare with PDCT (r=0.81, P<0.001). However, the iCare measurements were significantly and consistently lower than that obtained with GAT (ΔIOP=1.68±2.0, P=0.002, 95% CI: 0.7-2.6) and with PDCT (ΔIOP=1.61±2.5, P=0.01, 95% CI: 0.4-2.8). CCT, ECD, CCT, AXL, corneal curvature or astigmatism did not influence IOP measurement by any instrument.

Conclusions: IOP measurement with three different techniques (applanation, rebound and dynamic contour) showed good correlations, despite an increased corneal thickness following DSAEK. However, the iCare, which is based on a rebound tonometry showed significant lower IOP then the two other methods. This should be taken into account when evaluating patients post DSAEK.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jfo.2016.09.009DOI Listing

Publication Analysis

Top Keywords

intraocular pressure
8
dynamic contour
8
contour tonometry
8
pressure measurement
4
measurement dsaek
4
dsaek icare
4
icare goldmann
4
goldmann applanation
4
applanation dynamic
4
tonometry
4

Similar Publications

Purpose: To report a case of corneoscleral juvenile xanthogranuloma (JXG) with progressive anterior segment involvement refractory to topical steroids.

Observations: A 4-month-old male was referred for a new-onset subconjunctival lesion in the right eye. He was found to have a thickened, yellow corneoscleral lesion and hyphema, presumed to be ocular JXG.

View Article and Find Full Text PDF

Purpose: The purpose of this study was to evaluate the correlation between axial length (AL) and retinal oxygen dynamic parameters in adult patients.

Methods: This was an observational cross-sectional study with 79 Chinese adults with myopia aged 18 to 37 years. All participants underwent AL measurements, cycloplegic refraction, and other ophthalmic examinations.

View Article and Find Full Text PDF

Purpose: Vascular impairments, including reduced capillary density (CD), impaired autoregulation capacity (Reg), and elevated intraocular pressure (IOP), have been identified as significant contributors to glaucomatous disease. This study implemented a theoretical model to quantify the impact of these impairments on retinal blood flow and oxygenation as intraluminal pressure (Pa) is varied.

Methods: A theoretical model of the retinal vasculature was used to simulate reductions in CD by 10% (early glaucoma) and 30% to 50% (advanced glaucoma), a range in autoregulation capacity from 0% (totally impaired) to 100% (totally functional), and normal (15 mm Hg) and elevated (25 mm Hg) levels of IOP.

View Article and Find Full Text PDF

Tubulin Acetylation Enhances Microtubule Stability in Trabecular Meshwork Cells Under Mechanical Stress.

Invest Ophthalmol Vis Sci

January 2025

Department of Ophthalmology, Duke Eye Center, Duke University, Durham, North Carolina, United States.

Purpose: To study the roles of tubulin acetylation and cyclic mechanical stretch (CMS) in trabecular meshwork (TM) cells and their impact on outflow pathway physiology and pathology.

Methods: Primary TM cell cultures were subjected to CMS (8% elongation, 24 hours), and acetylated α-tubulin at lysine 40 (Ac-TUBA4) was assessed by western blotting and immunofluorescence. Enzymes regulating tubulin acetylation were identified via siRNA-mediated knockdowns of ATAT1, HDAC6, and SIRT2.

View Article and Find Full Text PDF

Purpose: To evaluate the effectiveness of the PAUL Glaucoma Implant (PGI) and the adjunctive impact of intraoperative Mitomycin C (MMC) on surgical outcomes in patients with neovascular glaucoma (NVG).

Methods: This retrospective, comparative study included NVG patients who underwent PGI implantation. The cohort was divided into two groups: Group 1 (PGI without MMC) and Group 2 (PGI with MMC).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!