Intravoxel incoherent motion diffusion-weighted MRI of the abdomen: The effect of fitting algorithms on the accuracy and reliability of the parameters.

J Magn Reson Imaging

Department of Radiology and Research Institute of Radiology, University of Ulsan, College of Medicine, Asan Medical Center, 86, Asanbyeongwon-gil, Songpa-gu, Seoul, Korea.

Published: June 2017

Purpose: To evaluate the influence of fitting methods on the accuracy and reliability of intravoxel incoherent motion (IVIM) parameters, with a particular emphasis on the constraint function.

Materials And Methods: Diffusion-weighted (DW) imaging data were analyzed using IVIM-based full-fitting (simultaneous fit of all parameters) and segmented-fitting (step-by-step fit of each parameter), each with and without the constraint function, to estimate the molecular diffusion coefficient (D ), perfusion fraction (f), and flow-related diffusion coefficient (D ). Computational simulations were performed at variable signal-to-noise ratios to evaluate the relative error (RE) and coefficient of variation (CV) of the estimated IVIM parameters. DW imaging of the abdomen was performed twice at 1.5 Tesla using nine b-values (0-900 s/mm ) in 12 health volunteers (6 men and 6 women; mean age: 30 years). The measurement repeatability of IVIM parameters in the liver and the pancreas was evaluated using the within-subject coefficient of variation ( CV).

Results: In simulations, full-fitting without the constraint function yielded the largest RE (P < 0.001 for D and f; P ≤ 0.044 for D ) and CV (P ≤ 0.033 for D and f; P ≤ 0.473 for D ) for IVIM parameters among all four algorithms. In volunteer imaging, full-fitting without the constraint function also resulted in the poorest repeatability for D ( CV, 17.12%-65.45%) and f ( CV, 19.35%-42.84%) in the liver and pancreas, while the other algorithms had similar repeatability values ( CV, 4.05%-11.99% for D and 9.65%-18.66% for f). Measurement repeatability of D ( CV, 29.52%-85.01%) was the poorest among the IVIM parameters.

Conclusion: For accurate and reliable measurement of IVIM parameters, segmented fitting or full-fitting with the constraint function should be used for IVIM-based analysis of DW imaging.

Level Of Evidence: 3 Technical Efficacy: Stage 2 J. MAGN. RESON. IMAGING 2017;45:1637-1647.

Download full-text PDF

Source
http://dx.doi.org/10.1002/jmri.25535DOI Listing

Publication Analysis

Top Keywords

ivim parameters
12
intravoxel incoherent
8
incoherent motion
8
accuracy reliability
8
constraint function
8
diffusion coefficient
8
coefficient variation
8
parameters
5
motion diffusion-weighted
4
diffusion-weighted mri
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!