Objective: The aim of this study was to describe previously unrecognised or under-recognised adverse events associated with Melody® valve implantation.

Background: In rare diseases and conditions, it is typically not feasible to conduct large-scale safety trials before drug or device approval. Therefore, post-market surveillance mechanisms are necessary to detect rare but potentially serious adverse events.

Methods: We reviewed the United States Food and Drug Administration's Manufacturer and User Facility Device Experience (MAUDE) database and conducted a structured literature review to evaluate adverse events associated with on- and off-label Melody® valve implantation. Adverse events were compared with those described in the prospective Investigational Device Exemption and Post-Market Approval Melody® transcatheter pulmonary valve trials.

Results: We identified 631 adverse events associated with "on-label" Melody® valve implants and 84 adverse events associated with "off-label" implants. The most frequent "on-label" adverse events were similar to those described in the prospective trials including stent fracture (n=210) and endocarditis (n=104). Previously unrecognised or under-recognised adverse events included stent fragment embolisation (n=5), device erosion (n=4), immediate post-implant severe valvar insufficiency (n=2), and late coronary compression (n=2 cases at 5 days and 3 months after implantation). Under-recognised adverse events associated with off-label implantation included early valve failure due to insufficiency when implanted in the tricuspid position (n=7) and embolisation with percutaneous implantation in the mitral position (n=5).

Conclusion: Post-market passive surveillance does not demonstrate a high frequency of previously unrecognised serious adverse events with "on-label" Melody® valve implantation. Further study is needed to evaluate safety of "off-label" uses.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5425318PMC
http://dx.doi.org/10.1017/S1047951116002092DOI Listing

Publication Analysis

Top Keywords

adverse events
40
events associated
24
melody® valve
20
valve implantation
12
under-recognised adverse
12
adverse
11
events
10
post-market surveillance
8
associated melody®
8
unrecognised under-recognised
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!