The polarization of the irradiance from several 1000 W FEL lamps was measured between 450 and 900 nm. These lamps are universally used as irradiance calibration standards in radiometric laboratories. The irradiance was polarized between 2.3% and 3.2%, with the polarization axis aligned with the coiled filament, nearly perpendicular to the lamp axis. We have presented a simple model of the filament that explains the degree of polarization and the plane of polarization, based on the polarized emissivity of tungsten, and gives an approximate value for this polarization. While the irradiance is polarized, this polarization does not significantly effect the polarization of the light when reflected from a Spectralon plaque (Labsphere, Inc.). The polarization of these lamps should be considered when these FEL lamps are used to characterize optical instruments, particularly grating spectrometers without polarization scrambling devices.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5548118PMC
http://dx.doi.org/10.1364/AO.55.008829DOI Listing

Publication Analysis

Top Keywords

fel lamps
12
polarization
10
polarization irradiance
8
irradiance polarized
8
lamps
5
polarization properties
4
properties fel
4
lamps applied
4
applied radiometric
4
radiometric calibration
4

Similar Publications

The long-term temporal stability of a spectrograph is one of the most important characteristics affecting the spectrograph's radiometric performance. For many applications, from monitoring ocean color and lunar irradiance to laboratory irradiance measurement standards, the stability of a spectrograph is a primary factor in the overall measurement uncertainty and therefore is the major criterion for the suitability of the spectrograph as an optical-scale transfer standard. Here we report a facility built for testing the long-term radiometric stability of commercial, fiber-coupled spectrographs.

View Article and Find Full Text PDF

Bulb-shaped field emission lamps (FELs) with a helical cathode filament were simulated and fabricated in this research. The light bulbs comprised a helical stainless steel filament cathode grown with carbon nano-coils (CNCs) and an Al anode deposited on the bottom hemisphere of a 60-mm-diameter glass bulb. White light was generated when the field-emitted electrons bombarded a layer of three-color phosphor coated on the anode.

View Article and Find Full Text PDF

The polarization of the irradiance from several 1000 W FEL lamps was measured between 450 and 900 nm. These lamps are universally used as irradiance calibration standards in radiometric laboratories. The irradiance was polarized between 2.

View Article and Find Full Text PDF

Fabrication and efficiency enhancement of tubal field emission lamps (FELs) using multi-walled carbon nanotubes (MWNTs) as the cathode field emitters were studied. The cathode filaments were prepared by eletrolessly plating a nickel (Ni) film on the cathode made of a 304 stainless steel wire dip-coated with MWNTs. The 304 wire was dip-coated with MWNTs and nano-sized Pd catalyst in a solution, and then eletrolessly plated with Ni to form an MWNT-embedded composite film.

View Article and Find Full Text PDF

Spectral irradiance model for tungsten halogen lamps in 340-850 nm wavelength range.

Appl Opt

February 2010

Helsinki University of Technology (TKK), Metrology Research Institute, P.O.B. 3000, FI-02015 TKK, Espoo, Finland.

We have developed a physical model for the spectral irradiance of 1 kW tungsten halogen incandescent lamps for the wavelength range 340-850 nm. The model consists of the Planck's radiation law, published values for the emissivity of tungsten, and a residual spectral correction function taking into account unknown factors of the lamp. The correction function was determined by measuring the spectra of a 1000 W, quartz-halogen, tungsten coiled filament (FEL) lamp at different temperatures.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!