Purpose: To evaluate the use of a self-expanding tract sealant device (BioSentry™) on the rates of pneumothorax and chest tube insertion after percutaneous lung biopsy.
Materials And Methods: In this retrospective study, we compared 318 patients who received BioSentry™ during percutaneous lung biopsy (treated group) with 1956 patients who did not (control group). Patient-, lesion-, and procedure-specific variables, and pneumothorax and chest tube insertion rates were recorded. To adjust for potential selection bias, patients in the treated group were matched 1:1 to patients in the control group using propensity score matching based on the above-mentioned variables. Patients were considered a match if the absolute difference in their propensity scores was ≤equal to 0.02.
Results: Before matching, the pneumothorax and chest tube rates were 24.5 and 13.1% in the control group, and 21.1 and 8.5% in the treated group, respectively. Using propensity scores, a match was found for 317 patients in the treatment group. Chi-square contingency matched pair analysis showed the treated group had significantly lower pneumothorax (20.8 vs. 32.8%; p = 0.001) and chest tube (8.2 vs. 20.8%; p < 0.0001) rates compared to the control group. Sub-analysis including only faculty who had >30 cases of both treatment and control cases demonstrated similar findings: the treated group had significantly lower pneumothorax (17.6 vs. 30.2%; p = 0.002) and chest tube (7.2 vs. 18%; p = 0.001) rates.
Conclusions: The self-expanding tract sealant device significantly reduced the pneumothorax rate, and more importantly, the chest tube placement rate after percutaneous lung biopsy.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5215969 | PMC |
http://dx.doi.org/10.1007/s00270-016-1489-9 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!