Radiotherapy can affect the electronic components of a cardiac implantable electronic device (CIED) resulting in malfunction and/or damage. We sought to assess the incidence, predictors, and clinical impact of CIED dysfunction (CIED-D) after radiotherapy for cancer treatment. Clinical characteristics, cancer, different types of CIEDs, and radiation dose were evaluated. The investigation identified 230 patients, mean age 78 ± 8 years and 70% were men. A total of 199 patients had pacemakers (59% dual chamber), 21 (9%) cardioverter-defibrillators, and 10 (4%) resynchronizators or defibrillators. The left pectoral (n = 192, 83%) was the most common CIED location. Sixteen patients (7%) experienced 18 events of CIED-D after radiotherapy. Reset to backup pacing mode was the most common encountered dysfunction, and only 1 (6%) patient of those with CIED-D experienced symptoms of atrioventricular dyssynchrony. Those who had CIED-D tended to have a shorter device age at the time of radiotherapy compared to those who did not (2.5 ± 1.5 vs 3.8 ± 3.4 years, p = 0.09). The total dose prescribed to the tumor was significantly greater among those who had CIED-D (66 ± 30 vs 42 ± 23 Gy, p <0.0001). Multivariate logistic regression analysis identified the total dose prescribed to the tumor as the only independent predictor for CIED-D (odds ratio 1.19 for each increase in 5 Gy, 95% confidence interval 1.08 to 1.31, p = 0.0005). In conclusion, in this large population of patients with CIEDs undergoing radiotherapy for cancer treatment, the occurrence of newly diagnosed CIED-D was 7%, and the reset to backup pacing mode was the most common encountered dysfunction. The total dose prescribed to the tumor was a predictor of CIED-D. Importantly, although the unpredictability of CIEDs under radiotherapy is still an issue, none of our patients experienced significant symptoms, life-threatening arrhythmias, or conduction disorders.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.amjcard.2016.09.036DOI Listing

Publication Analysis

Top Keywords

cardiac implantable
8
implantable electronic
8
electronic device
8
cied-d radiotherapy
8
cied-d
5
radiotherapy-induced cardiac
4
device dysfunction
4
patients
4
dysfunction patients
4
patients cancer
4

Similar Publications

Use of Brain Death Recipients in Xenotransplantation: A Double-Edged Sword.

Xenotransplantation

January 2025

Beijing Key Laboratory of Preclinical Research and Evaluation for Cardiovascular Implant Materials, Animal Experimental Centre, National Centre for Cardiovascular Disease, Department of Cardiac Surgery, Fuwai Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China.

Organ transplants are used to treat many end-stage diseases, but a shortage of donors means many patients cannot be treated. Xenogeneic organs have become an important part of filling the donor gap. Many current studies of kidney, heart, and liver xenotransplantation have used gene-edited pig organs on brain-dead recipients.

View Article and Find Full Text PDF

Blood coagulation monitoring of patients with mechanical heart valve visiting Imam.

J Cardiothorac Surg

January 2025

Cardiovascular Research Center, Health Policy and Promotion Institute, Imam-Ali Hospital, Kermanshah University of Medical Sciences, Kermanshah, Iran.

Background: The lacking proper regulation of International Normalized Ratio (INR) as the main problem in patients with mechanical valve replacement surgery poses the risk of thrombosis and embolism on the one hand and the risk of bleeding on the other. For this reason, the correct monitoring of INR via the Time in Therapeutic Range (TTR) is needed. The present study aimed to explore the blood coagulation monitoring of patients with mechanical heart valve in Imam Ali Hospital of Kermanshah in 2021.

View Article and Find Full Text PDF

Risk factors for long-term severe tricuspid regurgitation following mitral valve replacement: a retrospective study.

BMC Cardiovasc Disord

January 2025

Department of Cardiology, Xuzhou Central Hospital, No.199 Jiefang South Road, Quanshan District, Xuzhou, 221009, People's Republic of China.

Background: The aim of this study is to identify factors associated with the development of long-term severe tricuspid regurgitation (TR) following mitral valve replacement (MVR).

Methods: A retrospective analysis was conducted involving 308 patients who underwent single-valve MVR at Xuzhou Central Hospital between April 2017 and December 2022. Preoperative color Doppler ultrasound indicated that all patients had either no or mild to moderate tricuspid regurgitation.

View Article and Find Full Text PDF

Background: Vascular thymus transplantation has been explored in animal models but remains untested in humans. Current approaches to congenital athymia involve avascular transplantation of allogeneic thymic tissue, which may delay immune recovery. Building on animal studies, we propose revascularization of thymic tissue in a human model.

View Article and Find Full Text PDF

The New Niche: Transcatheter Aortic Valve Explantation and the Key Issues Concerning the Heart Team.

Can J Cardiol

January 2025

Division of Cardiac Surgery, Rady Faculty of Health Sciences, University of Manitoba, Winnipeg, MB. Electronic address:

The volume of transcatheter aortic valve implantation (TAVI) for treatment of isolated aortic stenosis has far surpassed surgical aortic valve replacement (SAVR). There has been a consequent increase in TAVI explantation, now the fastest growing cardiac surgical procedure. Transcatheter heart valve explantation can be technically complex, with higher perioperative morbidity and mortality than routine SAVR or valve-in-valve TAVI.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!