Effect of Implant Diameter on Fatigue Strength.

Implant Dent

*Bachelor of Health Science, Department of Dental Laboratory Science and Engineering, Korea University, Seoul, Republic of Korea; Institute for Clinical Dental Research, Korea University Medical Center, Seoul, Republic of Korea. †Clinical Professor, Department of Advanced Prosthodontics, Institute for Clinical Dental Research, Korea University Guro Hospital, Seoul, Republic of Korea; Doctor of Dental Medicine, Kyung Hee University, Seoul, Republic of Korea. ‡Professor, Department of Advanced Prosthodontics, Institute for Clinical Dental Research, Korea University Guro Hospital, Seoul, Republic of Korea; Doctor of Medicine, Korea University, Seoul, Republic of Korea.

Published: February 2017

Purpose: To investigate the effect of implant diameter on fatigue strength using static and cyclic load test.

Materials And Methods: Four different implant systems-SuperLine (Φ4.0), NRLine (Φ3.1), SlimLine (Φ2.8, Φ2.3), and (Dentium)-were grouped by implant diameter. A static load test was conducted with 5 specimens for each group using a universal testing machine to measure the ultimate failure load (UFL). With 80% of the UFL in the weakest group, the starting load for a cyclic load test was determined and the test was performed with 8 specimens for each group. All tests were conducted according to ISO14801 (2007) until implant failure occurred. After dynamically loaded, each specimen was sectioned and stereo-microscopically examined. The failure modes of each implant system were classified. Static and cyclic load test data were respectively analyzed after the test of normality, with the level of significance at P = 0.05.

Results: In the static load test, the higher maximum load of the standard-diameter implant was significant compared with the recorded narrow or mini-implants (P < 0.05). The yield strengths of the Φ2.8 and Φ3.1 implants were significantly greater than that of the Φ2.3 implant (P < 0.05). In a cyclic load test, the mean number of cycles until implant failure occurred was recorded for each specimen. The value for the Φ4.0 implant was significantly greater (P < 0.001).

Conclusion: Implant diameter has an effect on the ability to withstand both static and cyclic loads within Dentium implant systems, The UFLs and fatigue cycles decreased as the implants diameter became smaller.

Download full-text PDF

Source
http://dx.doi.org/10.1097/ID.0000000000000502DOI Listing

Publication Analysis

Top Keywords

load test
20
implant diameter
16
cyclic load
16
implant
12
static cyclic
12
load
9
diameter fatigue
8
fatigue strength
8
static load
8
specimens group
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!