Introduction: Haemoglobin A (HbA) is widely used in the management of diabetes. Therefore, the reliability and comparability among different analytical methods for its detection have become very important.
Materials And Methods: A comparative evaluation of the analytical performances (precision, linearity, accuracy, method comparison, and interferences including bilirubin, triglyceride, cholesterol, labile HbA (LA), vitamin C, aspirin, fetal haemoglobin (HbF), and haemoglobin E (Hb E)) were performed on Capillarys 2 Flex Piercing (Capillarys 2FP) (Sebia, France), Tosoh HLC-723 G8 (Tosoh G8) (Tosoh, Japan), Premier Hb9210 (Trinity Biotech, Ireland) and Roche Cobas c501 (Roche c501) (Roche Diagnostics, Germany).
Results: A good precision was shown at both low and high HbA levels on all four systems, with all individual CVs below 2% (IFCC units) or 1.5% (NGSP units). Linearity analysis for each analyzer had achieved a good correlation coefficient (R > 0.99) over the entire range tested. The analytical bias of the four systems against the IFCC targets was less than ± 6% (NGSP units), indicating a good accuracy. Method comparison showed a great correlation and agreement between methods. Very high levels of triglycerides and cholesterol (≥ 15.28 and ≥ 8.72 mmol/L, respectively) led to falsely low HbA concentrations on Roche c501. Elevated HbF induced false HbA detection on Capillarys 2FP (> 10%), Tosoh G8 (> 30%), Premier Hb9210 (> 15%), and Roche c501 (> 5%). On Tosoh G8, HbE induced an extra peak on chromatogram, and significantly lower results were reported.
Conclusions: The four HbA methods commonly used with commercial analyzers showed a good reliability and comparability, although some interference may falsely alter the result.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5082223 | PMC |
http://dx.doi.org/10.11613/BM.2016.039 | DOI Listing |
Lab Med
November 2024
Department of Laboratory Medicine, Peking University Shenzhen Hospital, Shenzhen, China.
Background: Glycated hemoglobin, or hemoglobin A1c (HbA1c), serves as a crucial marker for diagnosing diabetes and monitoring its progression. We aimed to assess the interference posed by common Hb variants on popular HbA1c measurement systems.
Methods: A total of 63 variant and nonvariant samples with target values assigned by the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC)-calibrated methods were included.
Clin Chem Lab Med
September 2024
Department of Clinical Laboratory, 74639 Beijing Chao-Yang Hospital, Capital Medical University, Beijing, P.R. China.
Objectives: The impact of seven hemoglobin variants (Hb Q-Thailand, Hb G-Honolulu, Hb Ube-2, Hb New York, Hb J-Bangkok, Hb G-Coushatta, and Hb E) on the outcome of HbA was investigated for six methods by comparing with liquid chromatography-tandem mass spectrometry (LC/MS/MS) reference method.
Methods: Twenty-nine normal and 112 variant samples were measured by LC/MS/MS, Sebia Capillarys 3 TERA, Intelligene Biosystems QuanTOF, Premier Hb9210, Arkray HA-8190V, Bio-Rad D-100, and Tosoh G11, then evaluated for correlation, consistency, and mean relative bias among six methods. The lowest biological variation bias of ±2.
Ann Clin Lab Sci
January 2024
Department of Pathology and Laboratory Medicine, University of Kansas Medical Center, Kansas City, KS, USA
Scand J Clin Lab Invest
December 2023
Department of Clinical Biochemistry, Virgen del Rocío University Hospital, Seville, Spain.
Glycated hemoglobin measurements are a valuable tool for long-term blood glucose monitoring and the diagnosis of diabetes. Its widespread use has been made possible due to the development of new analytical methods with improved performances and standardization with reference materials. The aim of the present study was to evaluate the Trinity Biotech Premier Hb9210 analyzer for the measurement of HbA1c.
View Article and Find Full Text PDFScand J Clin Lab Invest
February 2023
Department of Laboratory Medicine, Shenzhen Longhua District Central Hospital, Shenzhen, China.
Hb variants prevalent in China are different from those in other countries. We aimed to assess the interference from Hb variants found in China on HbA1c measurement. All Hb variants were confirmed using Sanger sequencing.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!