Introduction: Haemoglobin A (HbA) is widely used in the management of diabetes. Therefore, the reliability and comparability among different analytical methods for its detection have become very important.

Materials And Methods: A comparative evaluation of the analytical performances (precision, linearity, accuracy, method comparison, and interferences including bilirubin, triglyceride, cholesterol, labile HbA (LA), vitamin C, aspirin, fetal haemoglobin (HbF), and haemoglobin E (Hb E)) were performed on Capillarys 2 Flex Piercing (Capillarys 2FP) (Sebia, France), Tosoh HLC-723 G8 (Tosoh G8) (Tosoh, Japan), Premier Hb9210 (Trinity Biotech, Ireland) and Roche Cobas c501 (Roche c501) (Roche Diagnostics, Germany).

Results: A good precision was shown at both low and high HbA levels on all four systems, with all individual CVs below 2% (IFCC units) or 1.5% (NGSP units). Linearity analysis for each analyzer had achieved a good correlation coefficient (R > 0.99) over the entire range tested. The analytical bias of the four systems against the IFCC targets was less than ± 6% (NGSP units), indicating a good accuracy. Method comparison showed a great correlation and agreement between methods. Very high levels of triglycerides and cholesterol (≥ 15.28 and ≥ 8.72 mmol/L, respectively) led to falsely low HbA concentrations on Roche c501. Elevated HbF induced false HbA detection on Capillarys 2FP (> 10%), Tosoh G8 (> 30%), Premier Hb9210 (> 15%), and Roche c501 (> 5%). On Tosoh G8, HbE induced an extra peak on chromatogram, and significantly lower results were reported.

Conclusions: The four HbA methods commonly used with commercial analyzers showed a good reliability and comparability, although some interference may falsely alter the result.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5082223PMC
http://dx.doi.org/10.11613/BM.2016.039DOI Listing

Publication Analysis

Top Keywords

premier hb9210
12
roche c501
12
comparative evaluation
8
evaluation analytical
8
analytical performances
8
capillarys flex
8
flex piercing
8
tosoh hlc-723
8
roche cobas
8
cobas c501
8

Similar Publications

Background: Glycated hemoglobin, or hemoglobin A1c (HbA1c), serves as a crucial marker for diagnosing diabetes and monitoring its progression. We aimed to assess the interference posed by common Hb variants on popular HbA1c measurement systems.

Methods: A total of 63 variant and nonvariant samples with target values assigned by the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC)-calibrated methods were included.

View Article and Find Full Text PDF

Evaluation of effects from hemoglobin variants on HbA measurements by different methods.

Clin Chem Lab Med

September 2024

Department of Clinical Laboratory, 74639 Beijing Chao-Yang Hospital, Capital Medical University, Beijing, P.R. China.

Objectives: The impact of seven hemoglobin variants (Hb Q-Thailand, Hb G-Honolulu, Hb Ube-2, Hb New York, Hb J-Bangkok, Hb G-Coushatta, and Hb E) on the outcome of HbA was investigated for six methods by comparing with liquid chromatography-tandem mass spectrometry (LC/MS/MS) reference method.

Methods: Twenty-nine normal and 112 variant samples were measured by LC/MS/MS, Sebia Capillarys 3 TERA, Intelligene Biosystems QuanTOF, Premier Hb9210, Arkray HA-8190V, Bio-Rad D-100, and Tosoh G11, then evaluated for correlation, consistency, and mean relative bias among six methods. The lowest biological variation bias of ±2.

View Article and Find Full Text PDF
Article Synopsis
  • The study compared HbA1c values from two analyzers: the Premier Hb9210 (HPLC method) and the Alinity c (enzymatic assay) during a lab transition.
  • Both analyzers demonstrated excellent precision in measuring HbA1c, with strong correlation in test results (r=0.99), indicating reliability.
  • However, Alinity c showed lower HbA1c values than the reference method at critical decision points, suggesting that results should be interpreted carefully and may require confirmation through HPLC.
View Article and Find Full Text PDF

Glycated hemoglobin measurements are a valuable tool for long-term blood glucose monitoring and the diagnosis of diabetes. Its widespread use has been made possible due to the development of new analytical methods with improved performances and standardization with reference materials. The aim of the present study was to evaluate the Trinity Biotech Premier Hb9210 analyzer for the measurement of HbA1c.

View Article and Find Full Text PDF

Hb variants prevalent in China are different from those in other countries. We aimed to assess the interference from Hb variants found in China on HbA1c measurement. All Hb variants were confirmed using Sanger sequencing.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!