Severity: Warning
Message: file_get_contents(https://...@gmail.com&api_key=61f08fa0b96a73de8c900d749fcb997acc09&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Aim: Ki67 is a prognostic and/or predictive biomarker in patients with malignancies. Flow cytometry is a powerful technology for single-cell multiparameter analysis.
Results: We developed and validated a multicolor quantitative flow cytometry assay for detection of intracellular Ki67 expression in various immune cell subsets from human blood. The assay was optimized and showed excellent precisions. Assessment of the sample stability indicated that percentage changes from the fresh sample for the reportable results of interest were within 20%, up to 72 h after blood collection in the Cyto-Chex BCT tube.
Conclusion: The validated assay is sufficiently robust to analyze clinical samples. Easy access to peripheral blood enables continuous monitoring of Ki67 expression in blood as a biomarker, for example, for immunotherapy studies.