Purpose: The primary objective of this work is to measure the secondary neutron field produced by an uncollimated proton pencil beam impinging on different tissue-equivalent phantom materials using organic scintillation detectors. Additionally, the Monte Carlo code mcnpx-PoliMi was used to simulate the detector response for comparison to the measured data. Comparison of the measured and simulated data will validate this approach for monitoring secondary neutron dose during proton therapy.
Methods: Proton beams of 155- and 200-MeV were used to irradiate a variety of phantom materials and secondary particles were detected using organic liquid scintillators. These detectors are sensitive to fast neutrons and gamma rays: pulse shape discrimination was used to classify each detected pulse as either a neutron or a gamma ray. The mcnpx-PoliMi code was used to simulate the secondary neutron field produced during proton irradiation of the same tissue-equivalent phantom materials.
Results: An experiment was performed at the Loma Linda University Medical Center proton therapy research beam line and corresponding models were created using the mcnpx-PoliMi code. The authors' analysis showed agreement between the simulations and the measurements. The simulated detector response can be used to validate the simulations of neutron and gamma doses on a particular beam line with or without a phantom.
Conclusions: The authors have demonstrated a method of monitoring the neutron component of the secondary radiation field produced by therapeutic protons. The method relies on direct detection of secondary neutrons and gamma rays using organic scintillation detectors. These detectors are sensitive over the full range of biologically relevant neutron energies above 0.5 MeV and allow effective discrimination between neutron and photon dose. Because the detector system is portable, the described system could be used in the future to evaluate secondary neutron and gamma doses on various clinical beam lines for commissioning and prospective data collection in pediatric patients treated with proton therapy.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1118/1.4963813 | DOI Listing |
Rev Sci Instrum
December 2024
Lawrence Livermore National Laboratory, Livermore, California 94550, USA.
Ultra-intense short-pulse lasers interacting with matter are capable of generating exceptionally bright secondary radiation sources. The short pulse duration (picoseconds to nanoseconds), small source size (sub-mm), and comparable high peak flux to conventional single particle sources make them an attractive source for radiography using a combination of particle species, known as multimodal imaging. Simultaneous x-ray and MeV neutron imaging of multi-material objects can yield unique advantages for material segmentation and identification within the full sample.
View Article and Find Full Text PDFJ Phys Chem B
December 2024
Center for Neutron Science, Department of Chemical and Biomolecular Engineering, College of Engineering, University of Delaware, Newark, Delaware 19711, United States.
Hydrogen-deuterium exchange (HDX) measured by small-angle neutron scattering (HDX-SANS) is used to measure HDX in bovine serum albumin (BSA) under different temperatures and formulation conditions. HDX-SANS measurements are performed at 40, 50, and 60 °C in DO after storing proteins at 4 °C for 1 week to pre-exchange the readily accessible hydrogens. This enables us to probe the long-time HDX of protons at the core of the BSA proteins, which is more challenging for solvent molecules to access.
View Article and Find Full Text PDFPhys Med
December 2024
Dosimetry for Radiation Therapy and Diagnostic Radiology, Physikalisch-Technische Bundesanstalt (PTB), Braunschweig 38116, Germany; Metrology Research Center, National Research Council of Canada, 1200 Montreal Road, Ottawa, ON, K1A0R6, Canada.
Background: FLASH radiotherapy necessitates the development of advanced Quality Assurance methods and detectors for accurate monitoring of the radiation field. This study introduces enhanced time-resolution detection systems and methods used to measure the delivered number of pulses, investigate temporal structure of individual pulses and dose-per-pulse (DPP) based on secondary radiation particles produced in the experimental room.
Methods: A 20 MeV electron beam generated from a linear accelerator (LINAC) was delivered to a water phantom.
Nat Commun
November 2024
PSI Center for Photon Sciences CPS, Villigen PSI, Switzerland.
Appl Radiat Isot
February 2025
Faculty of Physics, University of Science and Technology Houari Boumediene ( USTHB), BP 32 EL ALIA, BAB EZZOUAR, Algiers, Algeria.
The present study aimed to assess organ doses and the associated cancer risks related to secondary radiation (photons and neutrons) exposure during 3D Conformational Radiotherapy (3D-CRT) for patients with prostate cancer in Algeria. To this purpose, a detailed geometric Monte Carlo (MC) modeling of the LINAC, combined with a hybrid whole-body phantom was carried out. The secondary radiation doses were calculated in patient's organs, both within and outside the field.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!