Air-Water Binary Gas Integrity Test for Sterilizing and Virus Filters.

PDA J Pharm Sci Technol

MilliporeSigma, Bedford, MA.

Published: May 2018

Unlabelled: Reliability of retention performance is of paramount importance for membrane filters designed for sterile and virus filtration. To achieve dependable retention, an integrity test can be applied to ensure the absence of oversize pores or defects that can compromise the retention capability of the filter. Probably the most commonly applied nondestructive integrity test for membrane filters is the gas-liquid diffusion test, with air and water often used as the gas-liquid pair. However, the sensitivity of the air-water diffusion test is limited by the fact that the diffusive flow rate for an integral membrane can span a range that is large compared to the flow contributed by a defect. A novel nondestructive air-water integrity test for microporous and nanoporous membranes is introduced here that provides improved test sensitivity by measuring the gas composition in addition to gas flow rate. Oxygen permeates through water faster than does nitrogen, so with air as the challenge gas and water as the wetting fluid, the permeate stream will be enriched in oxygen. The permeate oxygen concentration is predictable, accurately measurable, and within a narrow and repeatable range for an integral membrane. A leak through the membrane will result in a deviation from the integral permeate concentration, signaling a defect. Compared to the conventional air-water diffusion test, this air binary gas (i.e., O and N) test in which the permeate gas composition is measured (in addition to the diffusive flow rate) has a superior signal-to-noise ratio and was demonstrated to provide a significantly higher level of retention assurance for both sterilizing grade and virus filters. Because air and water are used as the gas-liquid pair, the air binary gas test also maintains the convenience, safety, and environmentally friendly aspects of the air-water diffusion test.

Lay Abstract: To ensure that sterilizing and virus removal filters are free of defects, an integrity test is often conducted both before and after use of the filters. In the commonly used air-water diffusion integrity test, pressurized air is applied to the water wetted filter and the air flow rate across the filter is measured. A flow rate above a specified limit indicates a leak through the filter. The sensitivity of the test is limited by the level of background noise (integral flow rate) relative to the leak signal (excess flow rate). An enhancement to the air-water diffusion test is introduced here in which the sensitivity of the test can be improved by measuring the composition of the diffused gas. Oxygen permeates through water faster than does nitrogen, so the permeating gas will be enriched in oxygen. Compared to the flow rate, which can span a range of values for integral filters, the integral oxygen concentration is well defined, so even small deviations from the expected concentration signal a leak. Because air and water are used as the test materials, the developed approach achieved higher sensitivity without sacrificing the convenience, safety, and environmentally friendly aspects of the air-water diffusion test.

Download full-text PDF

Source
http://dx.doi.org/10.5731/pdajpst.2016.006437DOI Listing

Publication Analysis

Top Keywords

flow rate
32
integrity test
24
air-water diffusion
24
diffusion test
20
test
17
binary gas
12
air water
12
gas
9
flow
9
air-water
8

Similar Publications

Purpose: Atrial fibrillation (AF) is the most common chronic cardiac arrhythmia that increases the risk of stroke, primarily due to thrombus formation in the left atrial appendage (LAA). Left atrial appendage occlusion (LAAO) devices offer an alternative to oral anticoagulation for stroke prevention. However, the complex and variable anatomy of the LAA presents significant challenges to device design and deployment.

View Article and Find Full Text PDF

Analysis of the effect of platelet function and different doses of ticagrelor after flow diverter treatment of intracranial aneurysms.

Neurosurg Rev

January 2025

Neurosurgery Center, Department of Cerebrovascular Surgery, Engineering Technology Research Center of Education Ministry of China on Diagnosis and Treatment of Cerebrovascular Disease, Zhujiang Hospital, Southern Medical University, Guangzhou, China.

Ticagrelor has become the standard drug for the treatment of intracranial aneurysms (IAs) with flow diverters (FDs), but the dosage has not been standardized. The effect of platelet function on clinical and imaging prognosis remains unclear. This study aimed to show the effects of different doses of ticagrelor and platelet aggregation function on the clinical and imaging prognosis after FDs treatment of aneurysms.

View Article and Find Full Text PDF

Introduction: Altered vascular microcirculation is recognized as a risk factor for anastomotic leakage (AL) in colorectal surgery. However, few studies evaluated its impact on AL using different devices, with heterogeneous results. The present study reported the initial experience measuring gut microcirculatory density and flow with the aid of incidence dark-field (IDF) videomicroscopy (Cytocam, Braedius, Amsterdam, The Netherlands) comparing its operative outcome using a propensity score matching (PSM) model based on age, gender, and Charlson Comorbidity Index (CCI).

View Article and Find Full Text PDF

Development of nebulized inhalation delivery for fusion-inhibitory lipopeptides to protect non-human primates against Nipah-Bangladesh infection.

Antiviral Res

January 2025

CIRI, Centre International de Recherche en Infectiologie, INSERM U1111, CNRS, UMR5308, Univ Lyon, Université Claude Bernard Lyon 1, École Normale Supérieure de Lyon, 21 Avenue Tony Garnier, 69007 Lyon, France.

Nipah virus (NiV) is a lethal zoonotic paramyxovirus that can be transmitted from person to person through the respiratory route. There are currently no licensed vaccines or therapeutics. A lipopeptide-based fusion inhibitor was developed and previously evaluated for efficacy against the NiV-Malaysia strain.

View Article and Find Full Text PDF

Objective: Enhancing ventilatory effort during pulmonary function testing can help reveal flow limitations not evident in normal tidal breathing. This study aimed to assess the efficacy and tolerability of using a CO2/O2 gas mixture to enhance tidal breathing with a barometric whole-body plethysmography system in both healthy cats and those with feline lower airway disease (FLAD).

Methods: This prospective study included healthy cats and those with FLAD, which underwent pulmonary function testing and were exposed to a 10% CO2/90% O2 gas mixture in a barometric whole-body plethysmography chamber, with CO2 concentrations maintained within the target range of 5% to 10%.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!