Objective: To evaluate repeatability and reproducibility of tear osmolarity measured using the TearLab osmometer in normal dogs and to assess its diagnostic potential in dogs with keratoconjunctivitis sicca (KCS).

Animals Studied: Beagle dogs; six normal and five with KCS.

Procedures: Tear osmolarity and Schirmer tear test-1 (STT-1) values were obtained at various times. Normal dogs were assessed for diurnal variation and repeatability and reproducibility of measurements. Dogs with KCS were evaluated before and after 5 months' topical twice-daily therapy with 2% cyclosporine.

Results: Mean ± SD tear osmolarity (mOsm/L) was significantly higher in normal dogs (337.4 ± 16.2) than in dogs with KCS before therapy (306.2 ± 18.0; P < 0.0001), but not following therapy with 2% cyclosporine (330.5 ± 13.7; P = 1.00). Osmolarity readings lower than 325.5 mOsm/L were suggestive of KCS (84.8% sensitivity and 87.1% specificity). In normal dogs, tear osmolarity readings were stable during the daytime (P = 0.99). Repeated measurements revealed high variability and typically poor-to-moderate repeatability and reproducibility, although this was improved by taking three successive measurements at each session. Considering combined data from all dogs, a positive correlation existed between STT-1 and tear osmolarity measurements (Pearson's correlation test, P = 0.04, r = 0.62).

Conclusions: Canine tear osmolarity as determined by TearLab osmometer was variable, required multiple readings to be informative, and differed from values reported for humans. Dogs with KCS had a lower tear osmolarity than did normal dogs, and this increased following cyclosporine therapy.

Download full-text PDF

Source
http://dx.doi.org/10.1111/vop.12436DOI Listing

Publication Analysis

Top Keywords

tear osmolarity
28
normal dogs
24
dogs
13
tearlab osmometer
12
repeatability reproducibility
12
dogs kcs
12
osmolarity
9
osmometer normal
8
dogs keratoconjunctivitis
8
keratoconjunctivitis sicca
8

Similar Publications

To compare the long-term efficacy and safety of intense pulsed light (IPL) treatments using a 590-nm and an acne filter. In this prospective, randomized, paired-eye trial study, 30 patients with moderate and severe meibomian gland dysfunction (MGD) were followed up for at least one month after their last treatment. Group A received IPL treatment with an acne filter, a type of notch filter that blocks wavelengths between 600 and 800 nm, allowing IPL to emit wavelengths between 400-600 nm and 800-1200 nm.

View Article and Find Full Text PDF

Treatment of Meibomian Gland Dysfunction by Classical Eyelid Hygiene Measures With and Without Additional Lipid Substitution for Tear Film Stabilization.

Eye Contact Lens

December 2024

Department of Ophthalmology (I.W., B.S.), Saarland University Medical Center (UKS), Homburg/Saar, Germany; Methods Center (A.K.), Eberhard Karls University Tübingen, Germany; and Praxis Prof. Dr. Dieter Dausch (D.D.), Amberg, Germany.

Objective: This study aimed to document the treatment success of eyelid hygiene with liposomal suspension using new diagnostic tools and to determine whether additional lipid substitution provided measurable benefits in meibomian gland dysfunction.

Methods: A single-center controlled, partially single masked study was conducted. Group A used eyelid hygiene only, whereas group B additionally applied a liposomal eye spray.

View Article and Find Full Text PDF

Background: The enhancement of ocular drug delivery is achieved by using biopolymer carriers in liquid ophthalmic preparations. Using natural polysaccharides to formulate ocular medications is attractive because they are economical and readily available. The primary goal of the study was to assess the safety and effectiveness of Tamarind seed polysaccharide (TSP)/Hyaluronic acid (HA) containing Rohto eye drops (TSP/HA drops) in alleviating the signs and symptoms of dry eye disease (DED) under dry environmental conditions using controlled environmental chamber (CEC).

View Article and Find Full Text PDF

Influence of Lifestyle Factors on Ocular Surface Parameters in Relation to Age.

Life (Basel)

November 2024

GI-2092 Optometry, Departamento de Física Aplicada, Facultad de Óptica y Optometría, Universidade de Santiago de Compostela, Campus Vida s/n, 15701 Santiago de Compostela, Spain.

Article Synopsis
  • The study aimed to assess how Video Display Terminal use, contact lens wear, and eyedrop use impact eye discomfort among different age groups.
  • It involved 252 participants completing surveys on their eye health and lifestyle, with 185 symptomatic individuals receiving thorough ocular tests.
  • Findings indicated age influences ocular health, with older adults experiencing greater discomfort, and prolonged VDT use worsening eye symptoms, while no clear connections were found between contact lens wear and ocular parameters.
View Article and Find Full Text PDF
Article Synopsis
  • The paper discusses long-term outcomes of a novel surgical technique for reinnervating the lacrimal gland in patients with neurodeprivative dry eye disease, following an initial 1-year report.
  • A study involving 9 patients post-surgery highlights significant improvements in tear secretion and tear film health, with measured parameters showing notable enhancements at the final follow-up.
  • Patients reported high levels of satisfaction, although tear film osmolarity remained similar between the operated and non-operated eye.
View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!