Objectives: The objective was to investigate the acute mechanical effects of post-dilatation on bioresorbable scaffolds (BRS) as determined by optical coherence tomography (OCT).
Background: Post-dilatation with high-pressure balloons is regarded as a key component of BRS implantation for treatment of coronary artery stenoses. However, the impact of post-dilatation on BRS in vivo has not been thoroughly investigated.
Methods: OCT was performed after the implantation procedure of 51 everolimus-eluting or novolimus-eluting polylactic acid-based BRS with (n = 27) or without non-compliant balloon post-dilatation (n = 24). The number of malapposed struts, strut fractures, edge dissections, residual in-scaffold area stenosis, and incomplete scaffold apposition area was analyzed over the complete length of each BRS with a spacing of 1 mm.
Results: OCT revealed a significantly lower incomplete scaffold apposition area if post-dilatation was performed (0.16 ± 0.49 mm with post-dilatation vs. 2.65 ± 2.78 mm without post-dilatation, p < 0.001), as well as a significantly lower absolute number of malapposed struts (1 ± 2 with post-dilatation vs. 13 ± 13 without post-dilatation, p < 0.001). No significant differences regarding residual in-scaffold area stenosis, strut fracture, edge dissection, symmetry index, or eccentricity index were observed in patients with vs. without post-dilatation.
Conclusion: Post-dilatation of BRS with non-compliant balloons significantly reduces the number of malapposed struts and incomplete scaffold apposition area without inducing higher rates of edge dissection or strut fracture.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1007/s00392-016-1048-z | DOI Listing |
Eur Heart J Case Rep
January 2025
Department of Cardiology, Komaki City Hospital, 1-20 Joubushi, Komaki, Aichi 485-8520, Japan.
Background: Transcatheter aortic valve implantation (TAVI) is a safe and effective therapy for patients with severe aortic stenosis. A Stuck leaflet and severe intraprosthetic regurgitation after valve implantation occur rarely but can lead to sudden haemodynamic deterioration. We encountered a case of a stuck leaflet following post-dilatation with the Edwards Sapien 3 Ultra RESILIA valve.
View Article and Find Full Text PDFCatheter Cardiovasc Interv
January 2025
Heart Centre, Catharina Hospital, Eindhoven, The Netherlands.
Backgrounds: Post-dilatation after stenting with a non-compliant (NC) balloon can be used to improve overall percutaneous coronary intervention (PCI) result. Due to lack of evidence on the effect of post-dilatation on adverse clinical endpoints there is no consensus whether post-dilatation should be used routinely. The aim of the current study was to determine the contemporary practice of post-dilatation.
View Article and Find Full Text PDFCardiol J
October 2024
Department of Interventional Cardiology, Cardiovascular Hospital and Claude Bernard University, INSERM Unit 1060 CARMEN, Lyon, France.
Eur Heart J Case Rep
October 2024
GRN Hospital Weinheim, Department of Cardiology, Vascular Medicine & Pneumology, Röntgenstraße 1, 69469 Weinheim, Germany.
Int J Surg Case Rep
November 2024
Heart Center, Kaplan Medical Center, Rehovot, affiliated with the Hebrew University, Jerusalem, Israel. Electronic address:
Introduction And Importance: Transcatheter aortic valve implantation (TAVI) is a minimally invasive procedure used to treat severe aortic stenosis. While TAVI is generally safe and effective, it can be complicated by rare adverse events such as prosthetic leaflet dislodgment leading to acute free aortic regurgitation.
Case Presentation: We report the case of a female patient who experienced acute free aortic regurgitation following elective TAVI.