Background: The Quantra Hemostasis Analyzer is a novel diagnostic device that uses an ultrasound-based technology, sonic estimation of elasticity via resonance (SEER) sonorheometry, to characterize the dynamic changes in viscoelastic properties of a blood sample during coagulation. Cardiac surgery utilizing cardiopulmonary bypass (CPB) is associated with a significant impact on the coagulation system and can result in perioperative coagulopathy. The aim of this study was to correlate SEER sonorheometry results to corresponding rotational thromboelastometry (ROTEM) and laboratory parameters obtained before, during, and after CPB.

Methods: The Quantra uses a multiwell cartridge that performs 4 independent measurements with different combination of reagents. The output test results include Clot Time, Clot Stiffness, Fibrinogen and Platelet Contribution, Clot Time Ratio, and Heparinase Clot Time. Clot Time was compared with ROTEM INTEM clotting time and the adjusted partial thromboplastin time. Clot Stiffness was compared with ROTEM EXTEM. The Fibrinogen Contribution to the Clot Stiffness was correlated to ROTEM FIBTEM as well as fibrinogen concentration by the Clauss method. The Platelet Contribution to Clot Stiffness was compared with absolute platelet count and ROTEM-determined clot elasticity attributable to platelets.

Results: Fifty-five patients undergoing elective cardiac surgery were enrolled in this prospective observational study. Clot Time exhibited good correlation with ROTEM INTEM clotting time (pre-CPB r = 0.84, post-CPB r = 0.65) and adjusted partial thromboplastin time (pre-CPB r = 0.72, post-CPB r = 0.89); however, the majority of values were within a narrow normal range. Clot Stiffness exhibited significant correlation with ROTEM EXTEM A10 throughout the course of the study in all samples (r = 0.84). Fibrinogen Contribution correlated strongly with FIBTEM A10 (r = 0.85), and moderately with the fibrinogen concentration (r = 0.73) determined with the Clauss assay. The Platelet Contribution to Clot Stiffness showed moderate correlation to absolute platelet counts (r = 0.48). However, the correlation between Platelet Contribution and ROTEM-determined clot elasticity attributable to platelets was stronger (r = 0.78) than platelet number. All of the correlation coefficients were statistically significant with P < .001.

Conclusions: SEER sonorheometry demonstrates significant correlation with ROTEM for determining Clot Stiffness and assessing Fibrinogen Contribution. SEER sonorheometry results can provide valuable information about the coagulation status in patients undergoing cardiac surgery using CPB.

Download full-text PDF

Source
http://dx.doi.org/10.1213/ANE.0000000000001507DOI Listing

Publication Analysis

Top Keywords

clot stiffness
28
seer sonorheometry
20
clot time
20
cardiac surgery
16
platelet contribution
16
contribution clot
16
clot
14
time clot
12
fibrinogen contribution
12
correlation rotem
12

Similar Publications

Background:  Clot waveform analysis (CWA) provides a global assessment of hemostasis and may be useful for patients with cirrhosis with complex hemostatic abnormalities. This study aimed to assess the association between prothrombin time (PT-) and activated partial thromboplastin time (aPTT-) based CWA parameters and cirrhosis severity and prospectively evaluate the role of CWA in predicting mortality and acute decompensation (AD) over 1 year.

Methods:  This prospective study included adult patients with cirrhosis between June 2021 and December 2023 at Chulalongkorn University Hospital.

View Article and Find Full Text PDF

Objectives: To assess whether the Quantra-Qplus can provide the cutoff values for predicting transfusion thresholds after cardiopulmonary bypass.

Design: Prospective observational study.

Setting: Single-center university hospital.

View Article and Find Full Text PDF

Impact of thrombus composition on virtual thrombectomy procedures using human clot analogues mechanical data.

J Mech Behav Biomed Mater

January 2025

Department of Chemistry, Materials and Chemical Engineering 'Giulio Natta', Politecnico di Milano, Italy.

Endovascular thrombectomy (EVT) aims at restoring blood flow in case of acute ischemic stroke by removing the thrombus occluding a large cerebral artery. During the procedure with stent-retriever, the thrombus is captured within the device, which is then retrieved, subjecting the thrombus to several forces, potentially leading to its fragmentation. In silico studies, along with mechanical characterisation of thrombi, can enhance our understanding of the EVT, helping the development of new devices and interventional strategies.

View Article and Find Full Text PDF

Background: Chronic kidney disease (CKD) is on the rise, and over 50% of patients die from cardiac causes. Patients develop heart failure due to unelucidated reno-cardiac interactions, termed type 4 cardiorenal syndrome (CRS4). The aim of this study is to establish and characterize a reliable model of CRS4 in swine with marked cardiac diastolic dysfunction.

View Article and Find Full Text PDF
Article Synopsis
  • The spiny mouse exhibits the ability to heal wounds without scarring, which may be linked to unique features of its blood and clotting mechanisms.
  • Compared to Balb/c mice, spiny mice showed stronger blood clots, faster tail bleeding times, and higher levels of clottable fibrinogen, indicating superior hemostatic capabilities.
  • Histological analysis revealed that spiny mouse clots were densely packed with fibrin and had better plasma clot stiffness, suggesting that these characteristics could enhance their wound healing and regenerative abilities.
View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!