Diagnostic reference levels for digital mammography in New South Wales.

J Med Imaging Radiat Oncol

Faculty of Health Sciences, The University of Sydney, Lidcombe, New South Wales, Australia.

Published: February 2017

Introduction: This work aims to explore radiation doses delivered in screening mammography in Australia, with a focus on whether compressed breast thickness should be used as a guide when determining patient derived diagnostic reference levels (DRLs).

Methods: Anonymized mammograms (52,405) were retrieved from a central database, and DICOM headers were extracted using third party software. Women with breast implants, breast thicknesses outside 20-110 mm and images with incomplete exposure or quality assurance (QA) data were excluded. Exposure and QA information were utilized to calculate the mean glandular dose (MGD) for 45,054 mammograms from 61 units representing four manufacturers using previously well-established methods. The 75th and 95th percentiles were calculated across median image MGDs obtained for all included data and according to specific compressed breast thickness ranges.

Results: The overall median image MGD, minimum, maximum were: 1.39, 0.19 and 10.00 mGy, respectively, the 75th and 95th percentiles across all units' median image MGD for 60 ± 5 mm compressed breast thickness were 2.06 and 2.69 mGy respectively. Median MGDs, minimum, maximum, 75th and 95th percentiles were presented for nine compressed breast thickness ranges, DRLs for NSW are suggested for the compressed breast thickness range of 60 ± 5 mm for the whole study and three detector technologies CR, DR, and photon counting to be 2.06, 2.22, 2.04 and 0.79 mGy respectively.

Conclusion: MGD is dependent upon compressed breast thickness and it is recommended that DRL values should be specific to compressed breast thickness and image detector technology.

Download full-text PDF

Source
http://dx.doi.org/10.1111/1754-9485.12540DOI Listing

Publication Analysis

Top Keywords

compressed breast
28
breast thickness
28
75th 95th
12
95th percentiles
12
median image
12
breast
9
diagnostic reference
8
reference levels
8
specific compressed
8
image mgd
8

Similar Publications

Fat Grafting With Lymphedema Fat: From Trash to Treasure?

Plast Reconstr Surg Glob Open

January 2025

From the Department of Plastic and General Surgery, Turku University Hospital, Turku, Finland.

Liposuction is a common procedure for patients with lymphedema with nonpitting adipose tissue hypertrophy. However, routinely, the lipoaspirate is discarded. Experimental studies have shown that adipose-derived stem cells in fat may enhance the regenerative and lymphangiogenic effects of the fat.

View Article and Find Full Text PDF

Acute radiation dermatitis (ARD) is a prevalent adverse effect of radiotherapy in patients with breast cancer, and there is a lack of high-quality data regarding its prevention and management. This study employs a systematic and iterative process to compile the perspectives of Canadian radiation oncology, nursing, and dermatology experts, aiming to establish consensus-based recommendations for the prevention and management of ARD in breast cancer patients. A modified Delphi consensus was conducted with the participation of 19 experts from across Canada.

View Article and Find Full Text PDF

: This study compares system-reported organ doses (ODs) to manually calculated mean glandular doses (MGDs) in mammography across multiple centers and manufacturers in Dubai. : A retrospective study of 2754 anonymized mammograms from six clinics in Dubai were randomly retrieved from a central dose survey database. Organ doses were documented along with other dosimetry information like kVp, mAs, filter, target, compression force, and breast thickness.

View Article and Find Full Text PDF

Accelerated High-resolution T1- and T2-weighted Breast MRI with Deep Learning Super-resolution Reconstruction.

Acad Radiol

January 2025

Department of Diagnostic and Interventional Radiology, University Hospital Bonn, Venusberg-Campus 1, 53127 Bonn, Germany (N.M., C.L., A.S., A.I., T.D., L.B., D.K., C.C.P., A.L., J.A.L.).

Rationale And Objectives: To assess the performance of an industry-developed deep learning (DL) algorithm to reconstruct low-resolution Cartesian T1-weighted dynamic contrast-enhanced (T1w) and T2-weighted turbo-spin-echo (T2w) sequences and compare them to standard sequences.

Materials And Methods: Female patients with indications for breast MRI were included in this prospective study. The study protocol at 1.

View Article and Find Full Text PDF

Immunotherapy in the Battle Against Bone Metastases: Mechanisms and Emerging Treatments.

Pharmaceuticals (Basel)

November 2024

Department of Anatomy and Genetics, College of Medicine, Alfaisal University, Riyadh 11533, Saudi Arabia.

Bone metastases are a prevalent complication in advanced cancers, particularly in breast, prostate, and lung cancers, and are associated with severe skeletal-related events (SREs), including fractures, spinal cord compression, and debilitating pain. Conventional bone-targeted treatments like bisphosphonates and RANKL inhibitors (denosumab) reduce osteoclast-mediated bone resorption but do not directly impact tumor progression within the bone. This review focuses on examining the growing potential of immunotherapy in targeting the unique challenges posed by bone metastases.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!