Background: Despite intermittent control of tracheal cuff pressure (P ) using a manual manometer, cuff underinflation (<20 cmHO) and overinflation (>30 cmHO) frequently occur in intubated critically ill patients, resulting in increased risk of microaspiration and tracheal ischemic lesions. The primary objective of our study was to determine the efficiency of an electronic device in continuously controlling P . The secondary objective was to determine the impact of this device on the occurrence of microaspiration of gastric or oropharyngeal secretions.

Methods: Eighteen patients requiring mechanical ventilation were included in this prospective randomized controlled crossover study. They randomly received either continuous control of P with Mallinckrodt device for 24 h, followed by discontinuous control with a manual manometer for 24 h, or the reverse sequence. During the 48 h after randomization, P was continuously recorded, and pepsin and alpha amylase were quantitatively measured in tracheal aspirates. P target was 25 cmHO.

Results: Clinical characteristics were similar during the two study periods, as well as mean airway pressure. Percentage of time spent with cuff overinflation or underinflation was significantly lower during continuous control compared with routine care period [median (IQR) 0.8 (0.1, 2) vs 20.9 (3.1, 40.1), p = 0.0009]. No significant difference was found in pepsin [median (IQR) 230 (151, 300) vs 259 (134, 368), p = 0.95] or in alpha amylase level [median (IQR) 1475 (528, 10,333) vs 2400 (1342, 15,391), p = 0.19] between continuous control and routine care periods, respectively.

Conclusions: The electronic device is efficient in controlling P , compared with routine care using a manometer. Further studies are needed to evaluate the impact of this device on intubation-related complications. Trial registration ClinicalTrials.gov Identifier: NCT01965821.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5050178PMC
http://dx.doi.org/10.1186/s13613-016-0200-2DOI Listing

Publication Analysis

Top Keywords

electronic device
12
continuous control
12
routine care
12
[median iqr
12
efficiency electronic
8
tracheal cuff
8
cuff pressure
8
critically ill
8
ill patients
8
randomized controlled
8

Similar Publications

Humidity sensors have been widely used to monitor humidity in daily life, agriculture fields, and so on. However, conventional sensors are not suitable for wearable devices because of their large dimensions and rigid substrates. Hence, we report a fast response, highly sensitive, and fully flexible humidity sensor on a PI substrate based on the composite material of reduced graphene oxide (rGO)/MoS, with a response time of 0.

View Article and Find Full Text PDF

Background: Delayed lead perforation is a rare complication of cardiac implantable electronic device (CIED). Clinical presentations range from completely asymptomatic to pericardial tamponade. Surgical lead extraction is recommended and transvenous lead extraction (TLE) with surgical backup is an alternative method.

View Article and Find Full Text PDF

Colonoscopy is a valuable tool for colorectal cancer screening and health checkups, with increasing utilization annually. Assisted entry is a standard procedure during electronic colonoscopy. In China, most clinically assisted colonoscopy procedures involve a nurse directly applying abdominal pressure to the patient's abdomen.

View Article and Find Full Text PDF

We report a controlled deposition process using atmospheric plasma to fabricate silver nanoparticle (AgNP) structures on polydimethylsiloxane (PDMS) substrates, essential for stretchable electronic circuits in wearable devices. This technique ensures precise printing of conductive structures using nanoparticles as precursors, while the relationship between crystallinity and plasma treatment is established through X-ray diffraction (XRD) analysis. The XRD studies provide insights into the effects of plasma parameters on the structural integrity and adhesion of AgNP patterns, enhancing our understanding of substrate stretchability and bendability.

View Article and Find Full Text PDF

Human activity recognition (HAR) is one of the most important segments of technology advancement in applications of smart devices, healthcare systems & fitness. HAR uses details from wearable sensors that capture the way human beings move or engage with their surrounding. Several researchers have thus presented different ways of modeling human motion, and some have been as follows: Many researchers have presented different methods of modeling human movements.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!