Background: The aims of this study were to evaluate the intra- and inter-observer repeatability of intraocular pressure (IOP) estimates obtained using an iCare rebound tonometer and to investigate the variation in IOP measurements due to positional differences in probe placement on the cornea.

Methods: The IOP estimates were obtained by two observers, twice on one eye, of 114 subjects using an iCare tonometer on the central cornea. Additionally, IOP and corneal thickness were measured in 38 subjects at three locations, namely, central, at 1.5 mm nasally and at 1.5 mm temporally from the central cornea. Agreement among measurements was assessed using Bland and Altman plots and the difference in measurements obtained by the observers was compared using paired t-test. Values obtained from central, nasal and temporal regions were compared using one-way analysis of variance.

Results: The mean IOP measurements obtained by observer 1 on two attempts were 16.2 and 16.0 mmHg (p > 0.05) were significantly different from the IOP values obtained by observer 2 on two occasions (16.3 and 15.7 mmHg; p < 0.0001). The limits of agreement (LOA) of intra-observer repeatability were -2.9 to +2.6 mmHg and -3.4 to +2.2 for observers 1 and 2, respectively. The LOA of inter-observer repeatability for first and second sequences were -2.8 to +3.0 mmHg and -3.3 to +2.7 mmHg with the second sequence of measurements being significantly different (p = 0.03). Although the corneal thickness was significantly greater nasally and temporally, when compared to the central location, by 32 and 20 microns respectively (p < 0.0001), the measured IOPs on those locations were not significantly different, when compared to the central measurements (p = 0.14).

Conclusion: There is good intra- and inter-observer repeatability of IOP data, as obtained by the iCare rebound tonometer, although the results may be slightly influenced by practitioner experience. The difference in IOP estimates obtained by different observers is unlikely to be of clinical significance. Corneal thickness showed regional variation; however, this did not influence IOP measurements obtained from those locations.

Download full-text PDF

Source
http://dx.doi.org/10.1111/cxo.12454DOI Listing

Publication Analysis

Top Keywords

intraocular pressure
8
estimates icare
8
icare rebound
8
rebound tonometer
8
iop estimates
8
iop measurements
8
central cornea
8
iop
6
evaluation intraocular
4
pressure estimates
4

Similar Publications

Purpose: To present the case of a young patient with BRAF V600E-mutant cutaneous melanoma who developed bilateral choroidal metastases complicated by neovascular glaucoma (NVG) in both eyes following the interruption of nivolumab therapy.

Methods: A 28-year-old female with primary cutaneous melanoma of the left hand underwent surgical resection and adjuvant nivolumab. Immunotherapy was discontinued due to immune-related acute interstitial nephritis.

View Article and Find Full Text PDF

: The PreserFlo™ MicroShunt (PFMS) is a bleb-forming device considered to be less invasive than traditional glaucoma surgery such as trabeculectomy. This study evaluates the 1-year success rates as well as safety profile of PFMS in patients having high intraocular pressure (IOP) and/or glaucoma refractory to drop therapy with a history of prior intravitreal dexamethasone therapy. : A total of 16 eyes after PFMS implantation due to elevated IOP after intravitreal dexamethasone implant (DEX-I) administration were included in this retrospective cohort study.

View Article and Find Full Text PDF

: This study aimed to illustrate a novel method for improving presbyopia by drinking cassiae tea. : A total of 425 eyes from 425 participants (aged 52.5 ± 9.

View Article and Find Full Text PDF

The rising prevalence of myopia is a significant global health concern. Atropine eye drops are commonly used to slow myopia progression in children, but their long-term use raises concern about intraocular pressure (IOP). This study uses SHapley Additive exPlanations (SHAP) to improve the interpretability of machine learning (ML) model predicting end IOP, offering clinicians explainable insights for personalized patient management.

View Article and Find Full Text PDF

This study aimed to investigate the etiology, pathogens, antibiotic susceptibility, treatments, and factors influencing the visual prognosis of pediatric post-traumatic endophthalmitis (PTE) to provide valuable insights for clinical diagnosis and treatment. A total of 301 children were included, with 142 (47.2%) cultures yielding positive results.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!