Aim: To investigate whether an image acquisition stabilizer (IAS) mounted on the sidestream dark field camera (SDF) during gastrointestinal surgery improves image stability and acquisition.

Methods: Serosal SDF imaging was compared with SDF imaging combined with an IAS (SDF + IAS) during gastrointestinal surgery. Stability was assessed as the image drift in pixels and the time to obtain stable images. The success rate was determined as the percentage of analyzable images after recording. The effect of negative pressure from the IAS was determined during single-spot measurements and by comparing microvascular parameters between groups. Data are presented as mean ± SD.

Results: Sixty serosal measurements were performed per group; 87% were successful in the SDF group and 100% in the SDF + IAS group (p = 0.003). Image drift in the SDF group was 148 ± 36 versus 55 ± 15 pixels in the SDF + IAS group; p < 0.001. Time to stable image was 96 ± 60 s in the SDF group versus 57 ± 31 s in the SDF + IAS group; p = 0.03. No effect of negative pressure was seen.

Conclusion: The use of an IAS mounted on an SDF camera during serosal microvascular assessment improves the success rate of image acquisition and stability and reduces the time to stable image with no effect on the microcirculation.

Download full-text PDF

Source
http://dx.doi.org/10.1159/000448735DOI Listing

Publication Analysis

Top Keywords

sdf ias
16
image acquisition
12
gastrointestinal surgery
12
time stable
12
sdf group
12
ias group
12
sdf
11
image
8
acquisition stabilizer
8
sidestream dark
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!