Aims: Accurate placement of the acetabular component during total hip arthroplasty (THA) is an important factor in the success of the procedure. However, the reported accuracy varies greatly and is dependent upon whether free hand or navigated techniques are used. The aim of this study was to assess the accuracy of an instrument system that incorporates 3D printed, patient-specific guides designed to optimise the placement of the acetabular component.
Patients And Methods: A total of 100 consecutive patients were prospectively enrolled and the accuracy of placement of the acetabular component was measured using post-operative CT scans.
Patients And Methods: A total of 100 consecutive patients were prospectively enrolled and the accuracy of placement of the acetabular component was measured using post-operative CT scans.
Conclusion: Accurate placement of the acetabular component can be achieved using patient-specific guides and is superior to free hand techniques and comparable to navigated and robotic techniques. Cite this article: Bone Joint J 2016;98-B:1342-6.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1302/0301-620X.98B10.37808 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!