Background: HbA1c is used in the diagnosis and monitoring of diabetes mellitus (DM). Interference from hemoglobin variants is a well-described phenomenon, particularly with HPLC-based methods. While immunoassays may generate more reliable HbA1c results in the presence of some variants, these methods are susceptible to negative interference from high concentrations of HbF. We report a case where an accurate HbA1c result could not be obtained by any available method due to the presence of a compound hemoglobinopathy.
Methods: HbA1c was measured by HPLC, immunoassay, and capillary electrophoresis. Hemoglobinopathy investigation consisted of a CBC, hemoglobin fractionation by HPLC and electrophoresis, and molecular analysis.
Results: HbA1c analysis by HPLC and capillary electrophoresis gave no result. Analysis by immunoassay yielded HbA1c results of 5.9% (Siemens DCA 2000+) and 5.1% (Roche Integra), which were inconsistent with other markers of glycemic control. Hemoglobinopathy investigation showed HbC with the hereditary persistence of fetal hemoglobin-2 Ghana deletion.
Conclusion: Reliable HbA1c results may be unobtainable in the presence of some hemoglobinopathies. HPLC and capillary electrophoresis alerted the laboratory to the presence of an unusual hemoglobinopathy. Immunoassays generated falsely low results without warning, which could lead to missed diagnoses and under treatment of patients with DM.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.cca.2016.09.017 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!